Previous Article | Next Article ![]()
Appl Environ Microbiol. 1967 January; 15(1): 145-147
Copyright © 1967 American Society for Microbiology. All Rights Reserved.
The Upjohn Company, Kalamazoo, Michigan
ABSTRACT
An instrument has been developed which permits the automatic quantitation of the turbidities of microbiological assay samples. Assay tubes were fed to the instrument at the end of the incubation period. The turbidity readings were automatically converted to digital data which were printed on International Business Machines (IBM) cards and from which potencies were calculated by an IBM computer. The instrument operated at a speed of over 240 tube readings per hour and was totally automatic in sample-mixing, readout, and data recording. The instrument is being used routinely at The Upjohn Co. for the turbidimetric bioassay of vitamins, with a coefficient of variation among repeated turbidity readings of 0.12 to 0.23%.