AEM
Home Help [Feedback] [For Subscribers] [Archive] [Search] [Contents]
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by O'Brien, T. C.
Right arrow Articles by Tauraso, N. M.
Right arrow Search for Related Content
PubMed
Right arrow Articles by O'Brien, T. C.
Right arrow Articles by Tauraso, N. M.
Agricola
Right arrow Articles by O'Brien, T. C.
Right arrow Articles by Tauraso, N. M.

 Previous Article  |  Next Article 

Appl Environ Microbiol. 1971 February; 21(2): 311-315
Copyright © 1971 American Society for Microbiology. All Rights Reserved.

Statistical Evaluation of Diluents and Automatic Diluting and Pipetting Machines in Influenza Serology

Thomas C. O'Brien, Suresh Rastogi and Nicola M. Tauraso

Laboratory of Virology and Rickettsiology, and Biometrics Section, Division of Biologics Standards, National Institutes of Health, Bethesda, Maryland 20014

ABSTRACT

The use of three diluents (i.e., 0.01 M phosphate-buffered saline, PBS; PBS with 0.2% gelatin, PBS/GEL; and PBS with 0.4% bovine plasma albumin) and three methods (i.e., the standard tube macro-procedure, TUBE; the manual microtechnique, MANUAL; and the semiautomatic microtechnique, AUTO) were statistically compared for their reproducibility and sensitivity in determining hemagglutinin (HA) and hemagglutination-inhibition (HI) antibody titers. In the HA test, analyses of between-cell variances of the different methods showed the AUTO microtiter procedure to be more reproducible than the standard TUBE method. The MANUAL microtiter procedure was the least reproducible. In the HI test, the TUBE method was the most reproducible. No significant difference in the reproducibility of the diluents was observed in either the HA or HI test. When a comparison of the sensitivity of test methods and diluents was made for determining HA titers, the AUTO microtiter procedure and PBS/GEL diluent appeared to be the method and diluent of choice. Evaluation of another instrument, the autopipetter, which standardizes the volume of diluent to be added in the microtechnique, suggests that the reproducibility of the AUTO microtiter procedure might be further increased.


Appl Environ Microbiol. 1971 February; 21(2): 311-315
Copyright © 1971 American Society for Microbiology. All Rights Reserved.