Previous Article | Next Article ![]()
Appl Environ Microbiol. 1974 October; 28(4): 644-646
Copyright © 1974 American Society for Microbiology. All Rights Reserved.
St. Joseph's Hospital and Medical Center, Paterson, New Jersey 07503
ABSTRACT
Clinical blood cultures were collected in the Bactec 8A flask (Johnston Laboratories, Cockeysville, Md.) and examined macrosopically, microscopically, and radiometrically in an effort to determine which approach produced the fastest detection time. Of 360 blood cultures found to contain organisms by subculture, 334 were first detected by Bactec, 98 by macroscopic examination, and 68 by microscopic examination. Examination times were at 4, 8, 16, 24, 36, and 48 h after collection of the specimen. Sixteen hours after specimen collection, microscopic examination had detected 31 positive cultures, macroscopic examination had detected two positive cultures, and Bactec had detected 160 positive cultures. By the end of the first 24 h of incubation, Bactec had detected 313 (93%) of those cultures eventually found to be positive. Although Bactec produced the fastest detection time in an overwhelming majority of the cultures, it failed to detect three of three Candida spp. cultures, three of five Bacteroides spp. cultures, and six of 32 Enterococcus spp. cultures during the first 48 h of incubation.
| J. Bacteriol. | Microbiol. Mol. Biol. Rev. | Eukaryot. Cell | All ASM Journals |
|---|