Previous Article | Next Article 
Appl Environ Microbiol, July 1998, p. 2572-2577, Vol. 64, No. 7
0099-2240/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
Capillary Electrophoresis Measurements of
Electrophoretic Mobility for Colloidal Particles of Biological
Interest
J. R.
Glynn Jr.,
B. M.
Belongia,
R. G.
Arnold,
K. L.
Ogden, and
J. C.
Baygents*
Department of Chemical and Environmental
Engineering, The University of Arizona, Tucson, Arizona 85721
Received 12 December 1997/Accepted 17 April 1998
 |
ABSTRACT |
The electrophoretic mobilities of three bacterial strains were
investigated by capillary electrophoresis (CE) and were compared with
results obtained by microelectrophoresis (ME). The CE measurements yielded bimodal electropherograms for two of the strains, thus illustrating for the first time that surface charge variations within a monoclonal population can be probed by CE. Intrapopulation variations were not detected by ME. The mobilities of three chemically distinct types of latex microspheres were also measured. Differences between the mean mobilities obtained by CE and ME were not
statistically significant (P
0.50); the standard
deviations of the CE measurements were typically 2 to 10 times smaller
than those obtained by comparable ME measurements. The reproducibility
of CE permitted batch-to-batch mobility variations to be probed for the
bacteria (one of the strains exhibited such variations), and
aggregation was evident in one of the latex suspensions. These effects
were not measurable with ME.
 |
INTRODUCTION |
Over the past decade,
capillary electrophoresis (CE) has been developed into a
standard technique in biotechnology. CE is used for
analytical separation of charged solutes, such as carbohydrates, proteins, DNA, pharmaceuticals, herbicides, and pesticides, as well as lower-molecular-weight ions (1, 19, 27). CE is a
powerful clinical diagnostic tool for profiling, screening, and
detecting drugs, carbohydrates, lipids, enzymes, proteins, and nucleic
acids (46). Variants of CE, known as micellar
electrokinetic capillary chromatography (27) and
capillary electrochromatography (36), allow solutions
of nonionic analytes to be resolved.
Recently, CE has been applied to suspensions involving particles of
biological interest, including viruses (19), bacteria (9, 19), silica sols (31), and polystyrene
nanospheres (45). An advantage of electrophoretic
separation compared with filtration or centrifugation is that
electrophoresis is characteristically gentle and suitable for labile
compounds and microorganisms. For example, Ebersole and McCormick
(9) found that more than 90% of the bacteria that they
tested remained viable after CE, and other workers (13, 35)
have obtained similar viability results for cells separated
dielectrophoretically in electric fields having comparable strengths
(ca. 100 V/cm).
While the capabilities of CE to separate compounds and particles
are well-established (for reviews see references 2,
26, and 27), the method has not been
widely used to measure electrophoretic mobilities. Accurate
measurements of electrophoretic mobilities are important in many
biological and environmental sciences and technologies, ranging from
clinical diagnostics (40, 41) to biocolloid adhesion
(8, 10, 12, 17, 33, 39, 43, 44). Tobacco mosaic virus
(6, 15) and polymer latex mobilities (6, 15, 23)
have been characterized by CE, but Zhu and Chen (47) found
that for human erythrocytes, the electrophoretic mobility measured by
CE differed by 21% from previously published values. As there have
been no direct comparisons between CE and other techniques, we
determined electrophoretic mobilities by CE and by microelectrophoresis
(ME), and we report the results obtained here. Our objectives were
essentially to examine whether CE could be used to measure the
electrophoretic mobility of larger particles (diameter, ca. 1 µm) of
biological interest and to assess the accuracy of the CE method
compared with ME (8, 29, 42). We present results for the
electrophoretic mobilities of three strains of bacteria and three
chemically distinct microspheres (used in biocompatibility
determinations) at two pHs and three ionic strengths. The method that
we used was based on the work done with tobacco mosaic virus by
Grossman and Soane (15). Differences between CE and ME
mobility measurements were not statistically significant, and the CE
data typically showed lower variances than the ME data. In addition, it
was possible by using CE to detect electrophoretic heterogeneities in
two of the bacterial species, as well as aggregation in one of the
latex suspensions. To our knowledge, this was the first use of CE to
resolve mobility variations within a monoclonal population.
 |
MATERIALS AND METHODS |
Bacterial strains and preparation.
The bacterial strains
investigated in this study were A1264, a gram-negative, motile,
ellipsoidal bacterium (ratio of length to width, 2.5) that was isolated
from the Savannah River deep subsurface environment, and two strains
isolated from the Department of Energy site at Oyster, Va. The Oyster
groundwater isolates were CD1 (a gram-negative, nonmotile, ellipsoidal
bacterium with a ratio of length to width of 1.5) and PL2W31 (a
gram-positive, nonmotile, ellipsoidal bacterium with a ratio of length
to width of 2.5). The equivalent spherical radii of the bacteria used, based on their projected areas, were all approximately 0.5 µm. Cultures were grown to the early stationary phase (absorbance at 600 nm, approximately 1.5) in 250-ml Erlenmeyer flasks containing 100 ml of
10% PTYG (0.25 g of peptone per liter, 0.25 g of tryptone per
liter, 0.50 g of yeast extract per liter, 0.6 g of
MgSO4 · 7H2O per liter, 0.07 g of
CaCl2 · 2H2O per liter) at room
temperature (22 to 24°C) with constant agitation (150 rpm). The cells
were separated by centrifugation (20 min, 104 × g) and resuspended in 100 ml of filter-sterilized
3-(N-morpholino)propanesulfonic acid (MOPS) buffer
containing 4.186 g of MOPS (Sigma; titrated to pH 7.02 with 1 N NaOH;
conductivity, 760 µS/cm; ionic strength, 10 mM) per liter. The
washing procedure was repeated to produce each final suspension
(concentration, 109 to 1010 cells/ml), which
was then incubated at room temperature for 18 h. (The bacterial
species and stage of growth were predicated on the results of a related
bacterial attachment study (3); the CE technique which we
describe is robust and adaptable to species at all stages of growth.)
At that point, cell density was measured by acridine orange direct
counting (20). Throughout the process, culture purity was
repeatedly verified by direct visual observation (approximately
104 cells were observed), plating, and comparisons of
colony morphology. Direct visual observation was also used to confirm
that the cells were monodisperse at the time of the experiments.
CE measurements were performed with at least three separate batches of
each microorganism. The batches were grown and prepared for measurement
on different days. Typically, three to five mobility measurements were
obtained for a given batch, although in some instances, as few as one
or two measurements were obtained. ME measurements were obtained with
cells from a single culture.
Microspheres and preparation.
The following three types of
microspheres were studied: 0.6-µm-diameter, surfactant-free,
hydrophobic amidine polystyrene latex microspheres (Interfacial
Dynamics Corp., Portland, Oreg.); 0.22-µm-diameter, surfactant-free,
hydrophilic carboxylate-modified polystyrene latex (CML) microspheres
(Interfacial Dynamics Corp.); and 0.24-µm-diameter, hydrophobic
polymethyl methacrylate (PMMA) microspheres (NASA Space Science Lab,
Marshall Space Flight Center, Huntsville, Ala.). The hydrophobicities
reported below are the hydrophobicities under the test conditions used
in this study. The polystyrene and PMMA spheres were supplied at stock
concentrations equivalent to roughly 4 and 20% solids, respectively.
The stock solutions were centrifuged for 4 to 16 min, depending on the
size of the microspheres, resuspended in the appropriate carrier
buffers, and sonicated for at least 10 min to produce monodisperse
suspensions. The carrier buffers used were 100 mM borate (Beckman
borate buffer kit; pH 8.39; conductivity, 1.60 mS/cm) and 2 mM borate
(pH 8.35; conductivity, 41.3 µS/cm). The 2 mM borate buffer was
produced by diluting a 100 mM buffer solution with ultrapure water
(Ultra-Pure Water Research Group, Department of Chemical and
Environmental Engineering, The University of Arizona). The buffers were
filter sterilized. The washing procedure was repeated three times for the CML and amidine particles and 6 to 10 times for the PMMA particles since the PMMA stock solutions contained residual impurities from the
emulsion polymerization process. The suspensions were diluted to final
volume fractions of 0.01 to 0.001 and stored overnight at 4°C to
ensure that the spheres were in equilibrium with the carrier buffers.
The amidine and CML particle measurements were made in the 2 and 100 mM
borate buffers. The PMMA measurements were made in the 2 mM borate
buffer. Prior to measurements, the suspensions were inspected for
aggregation with a light microscope.
CE measurements.
A Beckman P/ACE System 2100 capillary zone
electrophoresis unit, controlled by Beckman System Gold software, was
used to measure electrophoretic mobilities. A schematic diagram of the
instrument is shown in Fig. 1.
Application of an electric field down the axis of the capillary drove
particle transport by the following two mechanisms: electrophoresis
(translation due to the particles' native charge) and electroosmosis
(bulk convection of the carrier electrolyte in which the particles were
immersed). M, the electrophoretic mobility of a particle
population (or subpopulation), follows from the translation rate
compared with the translation rate of a neutral dopant, which moves
from the inlet past the UV detector at the electroosmotic velocity of
the buffer (21, 28): M = (LtotLd/V)
(1/tp
1/tref), where
Ld is the distance from the inlet reservoir to
the detector, V is the voltage difference applied across the
length of the capillary (Ltot), and
tref and tp are the mean
migration times of the reference marker and the particle population (or
subpopulation), respectively (27).
The capillaries used were obtained from Polymicro Technologies
(Phoenix, Ariz.) and were flexible fused silica capillaries
(inside
diameter, 75 µm) that were not coated on the inside and
had an
exterior polyimide coating. The capillaries connected the
anode (inlet)
reservoirs to the cathode (outlet) reservoirs (Fig.
1) and were
typically 57 cm long (distance from the cathode reservoir
to the
detector, 50 cm). A detector window was created by burning
off the
polyimide coating 7 cm from the outlet; the detector cell
volume was
O(10
12 m
3). Prior to each use, the
capillaries were conditioned by using
the Beckman capillary replacement
procedure (a 10-min high-pressure
rinse with Beckman capillary
regenerator solution A, 5 min of
H
2O, 10 min of buffer).
The optimum detector wavelength was determined for each sample
suspension by measuring the absorbance of a capillary filled
with the
suspension over the range of possible wavelengths after
a baseline was
established for the background buffer. The optimum
detector wavelength
for all of the measurements was 214 nm. For
the bacteria, the lower
detection limit corresponded to an injected
bacterial concentration of
approximately 10
8 cells/ml (data not shown) or 2,500 cells
(for an injected volume
of 25 nl). For the microspheres, the detection
limit was 10
10 particles/ml.
Actual measurements were preceded by a prerinse with buffer (5 min for
bacteria and 2 min for latex microspheres). Samples
were introduced
into the capillary with a 2- to 10-s standard
high-pressure
(20-lb/in
2) injection. The injection time was a function of
the sample absorbance.
Typically, a 5-s injection was used, and for a
57-cm-long capillary
with an inside diameter of 75 µm the sample
volume injected was
approximately 25 nl. CD1 was also injected
electrokinetically
by using a 175-V/cm applied electric field for
10 s to introduce
the sample into the capillary. Heating effects
were avoided by
maintaining the capillary at 25°C during the
measurements and
by performing an Ohm's law test (
32) to
establish the operating
conditions for which there was a linear
relationship between the
applied electric field and the current within
the capillary. The
applied electric fields used ranged from 88 to 426 V/cm.
A neutral dopant, mesityl oxide (
15,
28), was added to each
sample (2 µl of mesityl oxide in 400 µl of suspension) to
determine
the electroosmotic velocity. Measurements were determined
for the
sample with and without the marker to verify that addition
of the
marker did not affect colloid migration. Electrokinetic
measurements
were also determined for the background buffer and
(for bacterial
samples) a bacterial filtrate, which was obtained
by passing the
bacterial suspension through a 0.2-µm-pore-size
Acrodisc sterile
filter. The measurements were determined at least
three times per
sample.
A series of high-pressure rinses (5 min with 0.1 N HCl, 5 min with
H
2O, 5 min with Beckman capillary regenerator solution
A,
and 5 min with H
2O for bacteria; 5 min with Beckman
capillary
regenerator solution A and 5 min with H
2O for
latex spheres) were
used between successive measurements. The contents
of the cathode
and anode buffer reservoirs were replaced after three
measurements
to minimize the effect of electrode reactions on buffer
composition
(
4). At the end of each period of
experimentation, the capillary
was prepared for storage by repeating
the rinse procedure twice
(for a total of three rinses). This standard
rinse procedure was
followed by successive 10-min high-pressure rinses
with methanol
and air, and then the capillary was stored dry.
ME measurements.
ME electrophoretic mobilities were measured
with a Lazer Zee model 501 meter (PenKem, Inc., Bedford Hills, N.Y.).
Measurements were made as recommended in the user's manual, at room
temperature. The model 501 meter did not have a thermostat and reported
the data as a
potential, which could be converted to
electrophoretic mobility by using the Smoluchowski equation
(37): M = (
0/µ)
, where µ is
the buffer viscosity,
is the relative permittivity of the buffer,
and
0 is the permittivity of a vacuum (8.854 × 10
12 C2/N · m2). The
parameter values internal to the model 501 meter were reference values
for water at 20°C. Measurements were repeated at least 10 times for
each particle.
 |
RESULTS |
Bacteria.
Capillary electropherograms for the carrier buffer
and the bacterial filtrate exhibited no electrophoretic peaks. The
strength of the electroosmotic flow (EOF) was 7.13 ± 0.55 µm · cm/V · s (n = 36). The marker did
not affect the magnitude or distribution of the electrophoretic
mobilities for the bacteria. Preliminary experiments showed that the
results were also independent of the injection technique employed.
Electropherograms of A1264 and CD1 exhibited two electrophoretic peaks,
as shown in Fig.
2. Neither A1264 nor CD1
exhibited
significant batch-to-batch variation in their electrophoretic
mobility modes, but some variation was observed in the fractions
of the
populations associated with these modes (Table
1). Note
that the mode mobilities,
M1 and
M2, varied little
from batch
to batch, and there was little variation in
Mavg, the area-weighted
average of
M1 and
M2. Similar
results were obtained for A1264
(data not shown).

View larger version (12K):
[in this window]
[in a new window]
|
FIG. 2.
Typical electropherograms for bacterial cells.
Measurements were obtained in 10 mM MOPS buffer with an injection
interval of 5 s. Unless indicated otherwise, an electric field of
175 V/cm was applied over a 57-cm column. (a) Neutral marker peak (2.2 min) and two bacterial peaks (2.8 and 3.3 min) for strain A1264. An
electric field of 426 V/cm was applied over a 47-cm column. (b) Neutral
marker peak (8.9 min) and two bacterial peaks (12.5 and 13.6 min) for
CD1. The column length was 77 cm. (c) Neutral marker peak (7.1 min) and
a single bacterial peak (7.7 min) for PL2W31.
|
|
The third bacterium tested, PL2W31, exhibited only one electrophoretic
peak (Fig.
2) and batch-to-batch variation in mobility.
The mobilities
of the batches were

0.51 ± 0.04 µm · cm/V
· s
(
n = 5),

0.45 ± 0.04 µm · cm/V · s (
n = 2), and

0.41 ± 0.02
µm · cm/V · s (
n = 3).
The CE and ME results are compared in Table
2. Differences in the mobilities obtained
with the two techniques are shown in
Table
2. The CE data for A1264 and
CD1 are reported as the mode
mobilities and relative peak areas,
averaged over all of the measurements;
the standard deviations shown
are the mean standard deviations
for the batches. No bimodal
distribution was detected by ME for
A1264 and CD1. Since
batch-to-batch variability was evident with
PL2W31, comparisons
between the measurement techniques were based
on results from a single
culture of this microorganism. CE and
ME measurements were not obtained
with cells from the same culture
for A1264 and CD1, as these organisms
showed no significant batch-to-batch
variability in
Mavg (Table
1).
Microspheres.
Light microscopy of the microspheres
revealed monodisperse suspensions except for the PMMA
particles, which exhibited significant aggregation. No
electrophoretic peaks were observed with the baseline. The
strength of the EOF was 10.77 ± 0.17 µm · cm/V · s (n = 9) in 2 mM borate buffer and
6.60 ± 0.18 µm · cm/V · s (n = 8)
in 100 mM borate buffer. The marker had no effect on the
magnitude or distribution of the electrophoretic mobilities for the
latex spheres. Typical electropherograms for amidine and CML
particles are shown in Fig. 3. The
results of the CE and ME experiments are summarized in Table 2.

View larger version (12K):
[in this window]
[in a new window]
|
FIG. 3.
Typical microsphere electropherograms.
Measurements were obtained in 100 mM borate buffer with an
applied electric field of 350 V/cm over a 57-cm column. (a) Neutral
marker peak (2.8 min) and a microsphere peak (3.5 min) for the amidine
polystyrene microspheres. The injection interval was 10 s. (b)
Neutral marker peak (2.8 min) and a microsphere peak (5.8 min) for the
CML microspheres. The injection interval was 1 s.
|
|
CE measurements obtained with the heterogeneous PMMA suspension
revealed a bimodal distribution in mobility (Fig.
4; Table
2). Aggregates of microspheres
were apparent when the suspension
was inspected by light microscopy.
After sonication for 1 h, the
PMMA aggregates were disrupted
(i.e., the suspension was monodisperse,
as determined by light
microscopy). CE measurements obtained for
the monodisperse suspension
yielded one electrophoretic peak (Fig.
4). Additional CE measurements
were determined after time intervals
ranging from 0 to 2.5 h to
test for reaggregation of the microspheres.
Time-dependent aggregation
was apparent as determined by both
visual inspection (light microscopy)
and the redevelopment of
a bimodal distribution in the electrophoretic
mobility data (Fig.
4).

View larger version (17K):
[in this window]
[in a new window]
|
FIG. 4.
Electropherograms for the PMMA suspension. Measurements
were obtained in 2 mM borate buffer with an applied electric field of
175 V/cm over a 57-cm column. The injection interval was 5 s. (a)
Neutral marker peak (4.2 min) and two microsphere peaks (7.0 and 7.7 min) for the polydisperse PMMA particles (prior to sonication). (b)
Evolution of bimodal electrophoretic mobility in PMMA suspensions.
Electropherograms based on suspensions were obtained immediately after
sonication (dashed line) and 2.5 h after sonication (solid
line).
|
|
 |
DISCUSSION |
The general agreement between the data obtained with the CE and ME
mobility measurement techniques is indicated in Fig.
5. Statistical differences in the mean
electrophoretic mobilities determined by CE and ME were investigated by
using a paired t test (P = 0.40)
(7), a sign test (P = 0.50), and a
Wilcoxon signed rank test (P = 0.47)
(14); differences between the means were
insignificant. Systematic differences associated with the choice
of colloid, pH, or ionic strength were not observed.

View larger version (21K):
[in this window]
[in a new window]
|
FIG. 5.
Comparison of the electrophoretic mobilities determined
by CE (MCE) and ME (MME). The error bars are 2 standard deviations wide. The diagonal is a 1:1 line representing
complete data agreement. The data are labeled according to sample
type.
|
|
The finding that CE results agree with ME results is not surprising.
Electrokinetic theory shows that, as long as the particle size is
no more than a few percent of the capillary diameter, the influence of
hydrodynamic interactions on mobility measurements is negligible
(25). Colloidal (e.g., van der Waals) and/or
hydrophobic interactions between the analyte and the capillary wall can
also bias mobility measurements. Such interactions are well-documented for proteins and other organic analytes (18, 24, 30), which tend to adsorb to quartz capillaries. Similar effects have been observed for 0.2-µm-diameter hydrophobic polystyrene and PMMA particles in capillaries with inside diameters of 25 µm
(6). Analyte-wall interactions can be diminished by using
capillaries with larger cross-sectional areas (5, 6), by
modifying the carrier buffer, or by using coated capillaries
(34). Coatings that reduce wall interactions, however, often
reduce EOF and increase analysis time.
Based on similar numbers of replicate measurements, the standard
deviations for CE results were typically much lower than the standard
deviations for ME measurements (Table 2). This was probably because a
single CE mobility measurement was based on (i) analysis of a
relatively large number of particles (approximately 104
particles) and (ii) the time required for the analyte to move distances
on the order of 0.1 m in an electric field stronger than 100 V/cm.
By contrast, ME and related measurement techniques (22)
detect particle translation over short distances (typically less than
100 µm) in weak fields (10 V/cm or less) and often rely on
observation of far fewer particles.
Two of the bacteria used, A1264 and CD1, and one of the microsphere
types used, PMMA, exhibited multimodal behavior during CE measurements.
Similar distributions were not obtained with ME. CE is a sensitive
analytical separation technique tailored to resolve mixtures of charged
solutes on the basis of their electrophoretic mobilities. The
sensitivity stems from the strength of the applied field and the
uniformity of the particle velocities over the capillary cross section.
Thus, when a large number of particles is sampled, an electrophoretic
histogram and a more detailed picture are obtained.
The results of related studies on bacterial sorption (3)
suggest that the bimodal electrophoretic mobilities of the CD1 and
A1264 populations were due to intrapopulation heterogeneities in
surface charge rather than aggregation, contamination, or artifactual peak splitting. Bimodal mobilities of cells have been observed in other
contexts, and in these cases the bimodality has been attributed to
variations in surface structure (the presence or absence of appendages
or capsular material) (8) and surface antigens
(38). Ebersole and McCormick (9) found that
differences in chain morphology can cause multimodal behavior in
bacterial electrophoretic mobility. Visual inspection confirmed that
the cells used here were monodisperse. Spread plating revealed no major
contaminant. The detection limit for bacteria was approximately 2,500 cells or roughly 1/10th the standard number of cells injected. The
presence of a significant contaminant (observable with CE as a separate
peak) should have been recognized by direct visual observation or by
differences in colony morphology following growth on solid media. The
multiple peaks were in no way due to the carrier buffer, cell exudates,
or mesityl oxide. Ermakov et al. (11) concluded (based on
both theory and experiments) that artifactual peak splitting is avoided
by buffering solutions at least 1 pH unit from the particle isoelectric
point. Since bacterial isoelectric points are generally in the range
from 2 to 4, depending on species and growth stage (16), and
since the pH of the MOPS buffer was 7.02, artifactual peak splitting is
an unlikely source of multimodal mobilities observed.
The PMMA microspheres also exhibited an electrophoretic mobility
distribution. Direct visual observation revealed both individual particles and aggregated clumps in suspensions used for the original CE
and ME measurements. Sonication subsequently produced a monodisperse suspension of PMMA microspheres that yielded a single peak in the
corresponding electropherograms. Subsequent CE measurements obtained with the same suspension revealed the time-dependent evolution
of a second peak. Within 2.5 h, the suspension was comprised of
mainly aggregated particles, as indicated by visual inspection. Over
the same period the area of the peak containing the monodisperse microspheres decreased, while the second peak grew, presumably due to
aggregation.
Despite the noticeable difference in the areas under the peaks in
their respective electropherograms, the numbers of amidine and CML
(Fig. 3) particles injected were actually similar. A likely explanation for this phenomenon is the differences in the specific absorbances of the particles, as the area of an electrophoretic peak is
a function of both the number of particles injected and the specific
absorbance of the particles. Since both colloid volume fractions (less
than 0.01 for both microorganisms and microspheres) and UV absorbance
(typically less than 0.1 absorbance unit) were low, detection artifacts
associated with high particle concentrations are unlikely.
In the CE measurements, electroosmotic flow was characterized by doping
the samples with a neutral marker, mesityl oxide. EOF measurements were
reproducible for all of the carrier buffers tested. The standard
deviation of the EOF in 10 mM MOPS buffer (pH 7.02) was roughly three
times the standard deviation of the values obtained at both ionic
strengths in the borate buffer (pH 8.35). This probably was due to the
neutral pH of MOPS solutions, since electroosmosis in uncoated
capillaries is extremely sensitive to pH changes in the range from pH 4 to 8 (27).
CD1 samples were injected both electrokinetically and by high pressure.
Electrokinetic injections have been known to bias samples
(27), since analytes with lower (more negative)
electrophoretic mobilities tend to remain in the sample vial. No
difference in either the distribution or the magnitude of the
electrophoretic mobilities exhibited by CD1 was observed in
response to the change in injection technique (data not shown).
CD1 exhibited multimodal behavior in electrophoretic mobility at
relatively low field strengths (175 V/cm). Higher applied electric
fields were required for A1264 because at low field strengths, microbial motility distorted the CE peaks. Attempts to uncover multimodal electrophoretic behavior in bacterial strain PL2W31 by
increasing the applied electric field to approximately 350 V/cm were
unsuccessful.
The reproducibility of the CE data allowed batch-to-batch
variations to be studied in the bacteria. A1264 and CD1
showed variations in the distributions of their populations between the
respective mobility modes, although this had only a minor effect on the
average mobilities of the populations. PL2W31 exhibited batch-to-batch variability in its (unimodal) mean electrophoretic mobility.
Investigators should be able to use CE to help assess the influence of
such variability on biophysical processes (e.g., attachment to
surfaces) or, alternatively, to help evaluate connections between
variations in surface charge density and outer membrane structure
and composition.
In summary, CE measurements of electrophoretic mobility were not
distinguishable from measurements of electrophoretic mobility obtained by ME. Based on similar numbers of replicate measurements, the
standard deviations for CE results were typically much lower than the
standard deviations for ME measurements. Distributed mobilities, which
were readily observed with CE for two bacterial species and one of the
polymer latices, could not be detected with ME.
 |
ACKNOWLEDGMENTS |
We acknowledge the following persons from The University of
Arizona: Wallace Clark of the Division of Biotechnology, Arizona Research Laboratories, for assistance with the CE instrument; and
Philip Haworth of the Department of Materials Science & Engineering for
assistance with the ME measurements. We also thank the following individuals for providing the particles used in this study: David Balkwill of Florida State University (A1264); Aaron Mills of the University of Virginia (PL2W31); Anthony Palumbo of Oak Ridge National
Laboratory (CD1); and Dale Kornfeld of the NASA Marshall Space Flight
Center (PMMA).
This work was supported in part by grant DE-FG03-94ER61887 from the
Department of Energy Subsurface Science Program and by a grant from The
University of Arizona Foundation and the Office of the Vice President
for Research.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Department of
Chemical and Environmental Engineering, The University of Arizona,
Tucson, AZ 85721. Phone: (520) 621-6043. Fax: (520) 621-6048. E-mail: jcb{at}maxwell.che.arizona.edu.
 |
REFERENCES |
| 1.
|
Altira, K. D.
1996.
Additional application areas of capillary electrophoresis.
Methods Mol. Biol.
52:309-344[Medline].
|
| 2.
|
Baker, D. R.
1995.
Capillary electrophoresis.
John Wiley and Sons, New York, N.Y.
|
| 3.
| Baygents, J. C., J. R. Glynn, Jr., O. Albinger, B. K. Biesemeyer, K. L. Ogden, and R. G. Arnold. Variation of surface charge density in monoclonal
bacterial populations: implications for transport through porous media.
Environ. Sci. Technol., in press.
|
| 4.
|
Bello, M. S.
1996.
Electrolytic modification of a buffer during capillary electrophoresis.
J. Chromatogr.
744:81-91.
|
| 5.
|
Bello, M. S.,
R. Rezzonico, and P. G. Righetti.
1994.
Use of Taylor-Aris dispersion for measurement of a solute diffusion coefficient in thin capillaries.
Science
266:773-776[Abstract/Free Full Text].
|
| 6.
|
Belongia, B. M., and J. C. Baygents.
1997.
Measurements on the diffusion coefficient of colloidal particles by Taylor-Aris dispersion.
J. Colloid Interface Sci.
195:19-31[Medline].
|
| 7.
|
Box, G. E. P.,
W. G. Hunter, and J. S. Hunter.
1978.
Statistics for experiments.
John Wiley and Sons, New York, N.Y.
|
| 8.
|
Cowan, M. M.,
H. C. van der Mei,
I. Stokroos, and H. J. Busscher.
1992.
Heterogeneity of surfaces of subgingival bacteria as detected by zeta potential measurements.
J. Dent. Res.
71:1803-1806[Abstract/Free Full Text].
|
| 9.
|
Ebersole, R. C., and R. M. McCormick.
1993.
Separation and isolation of viable bacteria by capillary zone electrophoresis.
Bio/Technology
11:1278-1282[Medline].
|
| 10.
|
Eisen, A., and G. Reid.
1989.
Effect of culture media on Lactobacillus hydrophobicity and electrophoretic mobility.
Microb. Ecol.
17:17-25.
|
| 11.
|
Ermakov, S. V.,
M. Y. Zhukov,
L. Capelli, and P. G. Righetti.
1994.
Experimental and theoretical study of artificial peak splitting in capillary electrophoresis.
Anal. Chem.
66:4034-4042.
|
| 12.
|
Gannon, J.,
Y. Tan,
P. Baveye, and M. Alexander.
1991.
Effect of sodium chloride on transport of bacteria in a saturated aquifer material.
Appl. Environ. Microbiol.
57:2497-2501[Abstract/Free Full Text].
|
| 13.
|
Gascoyne, P. R. C.,
X.-B. Wang,
Y. Huang, and F. F. Becker.
1997.
Dielectrophoretic separation of cancer cells from blood.
IEEE (Inst. Electr. Electron. Eng.) Trans. Ind. Appl.
33:670-678.
|
| 14.
|
Gibbons, J. D.
1971.
Nonparametric statistical inference.
McGraw-Hill Book Company, San Francisco, Calif.
|
| 15.
|
Grossman, P. D., and D. S. Soane.
1990.
Orientation effects on the electrophoretic mobility of rod-shaped molecules in free solution.
Anal. Chem.
62:1592-1596[Medline].
|
| 16.
|
Harden, V. P., and J. O. Harris.
1953.
The isoelectric point of bacterial cells.
J. Bacteriol.
65:198-202[Free Full Text].
|
| 17.
|
Heckels, J. E.,
B. Blackett,
J. S. Everson, and W. E. Ward.
1976.
The influence of surface charge on the attachment of Neisseria gonorrohoeae to human cells.
J. Gen. Microbiol.
96:359-364[Abstract/Free Full Text].
|
| 18.
|
Hjertén, S.
1985.
High-performance electrophoresis elimination of electroendosmosis and solute adsorption.
J. Chromatogr.
347:191-198.
|
| 19.
|
Hjertén, S.,
K. Elenbring,
F. Kilár,
J. Liao,
A. J. C. Chen,
C. J. Siebert, and M. Zhu.
1987.
Carrier-free zone electrophoresis, displacement electrophoresis and isoelectric focusing in a high-performance electrophoresis apparatus.
J. Chromatogr.
403:47-61[Medline].
|
| 20.
|
Hobbie, J. E.,
R. J. Daley, and S. Jasper.
1977.
Use of Nucleopore filters for counting bacteria by fluorescence microscopy.
Appl. Environ. Microbiol.
33:1225-1228[Abstract/Free Full Text].
|
| 21.
|
Hunter, R. J.
1981.
Zeta potential in colloid science.
Academic Press, San Francisco, Calif.
|
| 22.
|
Hunter, R. J.
1993.
Introduction to modern colloid science.
Oxford University Press, New York, N.Y.
|
| 23.
|
Jones, H. K., and N. E. Ballou.
1990.
Separations of chemically different particles by capillary electrophoresis.
Anal. Chem.
62:2484-2490.
|
| 24.
|
Jorgenson, J. W., and K. D. Lukacs.
1983.
Capillary electrophoresis.
Science
222:266-272[Free Full Text].
|
| 25.
|
Keh, H. J., and J. L. Anderson.
1985.
Boundary effects on electrophoretic motion of colloidal spheres.
J. Fluid Mech.
153:417-439.
|
| 26.
|
Kuhn, R., and S. Hoffstetter-Kuhn.
1993.
Capillary electrophoresis: principles and practice.
Springer-Verlag, New York, N.Y.
|
| 27.
|
Landers, J. P. (ed.).
1994.
Handbook of capillary electrophoresis.
CRC Press, Ann Arbor, Mich.
|
| 28.
|
Lauer, H. H., and D. McManigill.
1986.
Capillary zone electrophoresis of proteins in untreated fused silica tubing.
Anal. Chem.
58:166-170.
|
| 29.
|
Lemp, J. F., Jr.,
E. D. Asbury, and E. O. Ridenour.
1971.
Electrophoresis of colloidal biological particles.
Biotechnol. Bioeng.
8:17-47.
|
| 30.
|
Liu, J.,
V. Dolnik,
Y.-Z. Hiesh, and M. Novotny.
1992.
Experimental evaluation of the separation efficiency in capillary electrophoresis using open tubular and gel-filled columns.
Anal. Chem.
64:1328-1336[Medline].
|
| 31.
|
McCormick, R. M.
1991.
Characterization of silica sols using capillary zone electrophoresis.
J. Liq. Chromatogr.
14:939-952.
|
| 32.
|
Nelson, R. J.,
A. Paulus,
A. S. Cohen,
A. Guttman, and B. L. Krager.
1989.
Use of Peltier thermoelectric devices to control column temperature in high-performance capillary electrophoresis.
J. Chromatogr.
480:111-127.
|
| 33.
|
Olsson, J.,
P. O. Glantz, and B. Krasse.
1976.
Electrophoretic mobility of oral streptococci.
Arch. Oral Biol.
21:605-609[Medline].
|
| 34.
|
Palmieri, R., and J. A. Nolan.
1994.
Protein capillary electrophoresis: theoretical and experimental considerations for method development, p. 325-368.
In
J. P. Landers (ed.), Handbook of capillary electrophoresis. CRC Press, Ann Arbor, Mich.
|
| 35.
|
Pethig, R.
1996.
Dielectrophoresis: using inhomogeneous AC electric fields to separate and manipulate cells.
Crit. Rev. Biotechnol.
16:331-348.
|
| 36.
|
Ross, G.,
M. Dittman,
F. Bek, and G. Rozing.
1996.
Capillary electrochromatography: enhancement of LC separation in packed capillary columns by means of electrically driven mobile phases.
Am. Lab. (Fairfield Conn.)
28:34-38.
|
| 37.
|
Russel, W. B.,
D. A. Saville, and W. R. Schowalter.
1989.
Colloidal dispersions.
Cambridge University Press, New York, N.Y.
|
| 38.
|
Sabolovic, D.,
E. Knippel,
U. Thomaneck,
J. Rychly, and W. Shütt.
1985.
Electrophoretic mobility and monoclonal antibody combined in the study of human peripheral blood lymphocytes, p. 333-343.
In
W. Shütt, and H. Klinkmann (ed.), Cell electrophoresis. Walter de Gruyter, Inc., New York, N.Y.
|
| 39.
|
Scholl, M. A., and R. W. Harvey.
1992.
Laboratory investigations on the role of sediment surface and groundwater chemistry in transport of bacteria through a contaminated sandy aquifer.
Environ. Sci. Technol.
26:1410-1417.
|
| 40.
|
Schütt, W.,
N. Hashimoto, and M. Shimizu.
1994.
Application of cell electrophoresis for clinical diagnosis, p. 255-280.
In
J. Bauer (ed.), Cell electrophoresis. CRC Press, Ann Arbor, Mich.
|
| 41.
|
Uhlenbruck, G.,
A. Fröml,
R. Lütticken, and K. Hannig.
1988.
Cell electrophoresis of group B streptococci: separation of types Ia, Ib/c, III and IV before and after neuraminidase treatment.
Zentralbl. Bakteriol. Parasintenkd. Infektionskr. Hyg. Abt. 1 Reihe A
270:28-24.
|
| 42.
|
van der Mei, H. C.,
J. de Vries, and H. J. Busscher.
1993.
Hydrophobic and electrostatic cell surface properties of thermophilic dairy streptococci.
Appl. Environ. Microbiol.
59:4305-4312[Abstract/Free Full Text].
|
| 43.
|
van Loosdrecht, M. C. M.,
J. Lyklema,
W. Norde,
G. Schraa, and A. J. B. Zehnder.
1987.
Electrophoretic mobility and hydrophobicity as a measure to predict the initial steps of bacterial adhesion.
Appl. Environ. Microbiol.
53:1898-1901[Abstract/Free Full Text].
|
| 44.
|
van Loosdrecht, M. C. M.,
W. Norde,
J. Lyklema, and A. J. B. Zehnder.
1990.
Hydrophobic and electrostatic parameters in bacterial adhesion.
Aquat. Sci.
52:103-114.
|
| 45.
|
VanOrman, B. B., and G. L. McIntire.
1989.
Analytical separation of polystyrene nanospheres by capillary electrophoresis.
J. Microcolumn Sep.
1:289-293.
|
| 46.
|
Xu, Y.
1995.
Capillary electrophoresis.
Anal. Chem.
67:463R-473R[Medline].
|
| 47.
|
Zhu, A., and Y. Chen.
1989.
High-voltage capillary zone electrophoresis of red blood cells.
J. Chromatogr.
470:251-260[Medline].
|
Appl Environ Microbiol, July 1998, p. 2572-2577, Vol. 64, No. 7
0099-2240/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
van Merode, A. E. J., van der Mei, H. C., Busscher, H. J., Krom, B. P.
(2006). Influence of Culture Heterogeneity in Cell Surface Charge on Adhesion and Biofilm Formation by Enterococcus faecalis.. J. Bacteriol.
188: 2421-2426
[Abstract]
[Full Text]
-
van Merode, A. E. J., van der Mei, H. C., Busscher, H. J., Waar, K., Krom, B. P.
(2006). Enterococcus faecalis strains show culture heterogeneity in cell surface charge.. Microbiology
152: 807-814
[Abstract]
[Full Text]
-
Butkus, M. A., Bays, J. T., Labare, M. P.
(2003). Influence of Surface Characteristics on the Stability of Cryptosporidium parvum Oocysts. Appl. Environ. Microbiol.
69: 3819-3825
[Abstract]
[Full Text]
-
Desai, M. J., Armstrong, D. W.
(2003). Separation, Identification, and Characterization of Microorganisms by Capillary Electrophoresis. Microbiol. Mol. Biol. Rev.
67: 38-51
[Abstract]
[Full Text]
-
Streger, S. H., Vainberg, S., Dong, H., Hatzinger, P. B.
(2002). Enhancing Transport of Hydrogenophagaflava ENV735 for Bioaugmentation of Aquifers Contaminated with Methyl tert-Butyl Ether. Appl. Environ. Microbiol.
68: 5571-5579
[Abstract]
[Full Text]
-
Dong, H., Rothmel, R., Onstott, T. C., Fuller, M. E., DeFlaun, M. F., Streger, S. H., Dunlap, R., Fletcher, M.
(2002). Simultaneous Transport of Two Bacterial Strains in Intact Cores from Oyster, Virginia: Biological Effects and Numerical Modeling. Appl. Environ. Microbiol.
68: 2120-2132
[Abstract]
[Full Text]
-
van der Mei, H. C., Busscher, H. J.
(2001). Electrophoretic Mobility Distributions of Single-Strain Microbial Populations. Appl. Environ. Microbiol.
67: 491-494
[Full Text]