Enumeration of viruses by EFM and flow cytometry is based on using highly fluorescent nucleic acid dyes. In the initial methods the approach used for enumerating bacteria in natural samples (17) was modified by staining the virus particles with DAPI (4',6-diamidino-2-phenylindole) and enumerating them on filters by EFM (11, 21, 23). Subsequently, the approach was modified by replacing DAPI with Yo-Pro-1 (12). Although Yo-Pro-1 has significant advantages over DAPI in that the fluorescence is more stable and the fluorescence yield and the binding coefficient for nucleic acids are higher, it has the disadvantages of relatively long staining times and interference by aldehyde-based fixatives. Hence, Yo-Pro-1-stained slides need to be made with freshly collected samples. Nonetheless, the accuracy and precision of viral counts obtained by EFM are considerably greater than the accuracy and precision of viral counts obtained by electron microscopy, and the procedure can be done in the field, making it a desirable approach for estimating viral abundance (14, 23).
In many instances it is necessary to preserve samples in the field and return them to the laboratory for analysis of viral abundance; consequently, efforts were made to modify the original Yo-Pro-1 method so that it could be used with fixed samples. One suggested approach was to use microwave radiation to allow rapid Yo-Pro-1 staining of fixed samples (26), while another modification was to replace Yo-Pro-1 with SYBR Green I (16) or SYBR Gold (7). Unfortunately, the introduction of the alternative dyes was not accompanied by data on the effect of fixation and storage of samples on estimates of viral abundance or by a comparison of the accuracy of the counts to the accuracy of the Yo-Pro-1 method counts. Subsequently, Bettarel et al. (2) compared the original Yo-Pro-1 method (12) with the modified methods involving microwave treatment (26) and replacement of Yo-Pro-1 by SYBR Green I (16). Their conclusion was that the SYBR Green I modification produced less stable fluorescence and yielded lower estimates of viral abundance but that the original Yo-Pro-1 and microwave-modified approaches produced estimates that were not significantly different.
In our laboratory, EFM is routinely used to estimate viral abundance in cultures and natural samples, and over the years we have counted thousands of samples. Although we typically employed the original Yo-Pro-1 method, we also used the modified methods. However, in our hands, the results were inconsistent. In particular, we experimented with the SYBR Green I modification and became suspicious that glutaraldehyde fixation as recommended in the protocol was affecting our results. In particular, we noticed decreases in viral abundance over time when viral isolates were preserved with aldehydes and stored at 4°C.
Aldehyde fixatives are routinely used to preserve natural samples, which are then typically stored at 4°C in the dark before slides are prepared for EFM. The length of time between fixation of samples and preparation of slides can vary from minutes to months or even years. Several studies have shown, however, that the abundance of bacteria decreases when the bacteria are fixed with glutaraldehyde, formaldehyde, or Lugol's iodine and stored at 4°C (10, 15, 22). Workers have also reported decreases in the estimates of viral abundance in aldehyde-preserved water (13, 14, 26, 27) and sediment (8) samples, although the data are quite limited. A recent study of viral isolates and natural marine samples by flow cytometry indicated that storage of aldehyde-fixed samples decreased by as much as 90% in just a few hours (5). This finding led to the suggestion that viruses used for enumeration by flow cytometry should be fixed with 0.5% glutaraldehyde for 15 to 30 min and then flash frozen in liquid nitrogen before they are stored at 80°C.
In this study, our goal was to firmly establish protocols that allow accurate estimation of viral abundance in natural waters by EFM. Specifically, we wished to determine whether aldehyde fixation of samples was responsible for inconsistent estimates of viral abundance in SYBR Green I-stained samples and, if so, what the magnitude of the error was over time. Second, we wished to establish a satisfactory protocol for the preservation of field samples so that accurate estimates of viral abundance can be obtained by EFM.
|
|
|---|
EFM.
For all experiments viral abundance was determined by EFM by using an Olympus AX70 epifluorescence microscope with a wide-blue filter set (excitation at 450 to 480 nm, 515-nm cutoff) and a UPlanApo x100 objective. The water samples were collected and either processed immediately or fixed with 0.5% glutaraldehyde (electron microscopy [EM] grade) or with filtered (pore size, 0.1 µm) 2.0% formaldehyde. Subsamples were either filtered directly or diluted with filtered (pore size, 0.02 µm) MilliQ water to reduce the number of viruses per field when preparations were observed under the microscope for ease of counting. Slides were made by filtering 800 µl of whole or diluted samples, either fixed or unfixed, through 0.02-µm-pore-size AlO3 filters (Anodisc; Whatman). The filters were immediately stained with SYBR Green I, SYBR Gold, or Yo-Pro-1 (Molecular Probes). SYBR Green I slides were prepared by using a modified method of Noble and Fuhrman (16). The filters were stained sample side up for a minimum of 15 min in the dark with diluted SYBR Green I (a 4 x 102 dilution in filtered [pore size, 0.02 µm] MilliQ water) in a plastic petri dish. Each filter was then mounted on a glass slide with 0.1% p-phenylenediamine (made freshly from a frozen 10% aqueous stock; Sigma Chemical Co.) in 50% glycerol-50% phosphate-buffered saline (0.05 M Na2HPO4, 0.85% NaCl; pH 7.5). The SYBR Gold slides were prepared by using the same method that was used for the SYBR Green I slides (7). Yo-Pro-1 slides were prepared by the method of Hennes and Suttle (12) for unfixed samples only. The filters were stained sample side up for 48 h in 100 µl of 50 mM Yo-Pro-1 in a 2 mM NaCN solution in a petri dish. The filters were then rinsed twice, with filtration, with 800 µl of filtered (pore size, 0.02 µm) MilliQ water and mounted onto glass slides with glycerol. The slides were either frozen or counted immediately by EFM. For frozen-storage experiments, the slides were stored at 20°C in a frost-free freezer until they were counted. In all cases, a minimum of 200 virus-like particles in at least 20 random fields were counted (19).
Preliminary experiments.
A series of preliminary experiments were conducted to determine if the effects of aldehyde preservation were dependent on the concentration of glutaraldehyde used for fixation. Duplicate samples were fixed in 0.5, 2, and 4% glutaraldehyde and stored at 4°C. Slides were prepared after 30 min and 1 and 2 days and then every 2 days until 12 days. Slides were counted immediately following preparation.
Glutaraldehyde versus formaldehyde.
To examine differences in preservation between glutaraldehyde and formaldehyde, a seawater sample collected from Jericho Beach was divided into two sets of triplicate subsamples. One set of subsamples was fixed in 0.5% glutaraldehyde (EM grade), and the other set was fixed in filtered 2.0% formaldehyde before the subsamples were incubated at 4°C in the dark for 30 min. Eight slides were then prepared from each subsample by using SYBR Green I as described above. One slide from each set was counted immediately, while the remaining slides were kept in a frost-free freezer at 20°C. On days 1, 2, 4, 7, 9, 13, and 16, one slide was prepared from each of the subsamples, which remained at 4°C in the dark, and immediately counted. One previously prepared frozen slide from each set was counted on days 1, 2, 7, 13 and 16 to compare the effects of storage of fixed samples at 4°C with the effects of storage of frozen slides on estimates of viral abundance. To examine short-term changes in viral abundance in fixed samples, this experiment was repeated, and slides for viral enumeration were prepared at 0, 1, 2, 4, 7, 12, 25, and 48 h.
Comparison of nucleic acid stains and storage methods.
Water samples were collected and divided into three sets of triplicate subsamples (designated sets I, II, and III) (Fig. 1). One set (set I) was left unfixed; the other two sets were fixed in 0.5% EM-grade glutaraldehyde (set II) or 0.74% filtered (pore size, 0.1 µm) formaldehyde (2% formalin) (set III) and stored at 4°C in the dark for 30 min. Immediately after fixation a slide was prepared from each triplicate subsample and stained with SYBR Green I or SYBR Gold. One Yo-Pro-1-stained slide was also made with each subsample of the unfixed triplicate subsamples. All the slides were counted on the same day by EFM. At least 400 virus-like particles were counted in 48 fields for this experiment. While the slides were being made, three 5-ml subsamples (subsamples A, B, and C) were taken from each treatment. Subsample A was flash frozen in liquid nitrogen and stored at 86°C by following the recommended storage procedures for flow cytometry (6), Subsample B was flash frozen and stored at 20°C in a non-frost-free freezer, and subsample C was frozen at 20°C. After 7 days, the subsamples were thawed in a 37°C water bath, and slides stained with SYBR Green I or SYBR Gold were made. At the same time, Yo-Pro-1-stained slides were made from the unfixed set of subsamples. The slides were kept frozen at 20°C in a frost-free freezer and counted within 7 days (Fig. 1). Samples were also stored at 4°C in the dark and sampled after 7 days.
![]() View larger version (19K): [in a new window] |
FIG. 1. Flow chart of comparison of nucleic acid stains and storage experiment. Natural water samples were collected in plastic bottles that were rinsed with water from collection sites. The samples were dispensed into three sets of triplicate containers and either left untreated (set I) or fixed with 0.5% glutaraldehyde (set II) or 0.74% formaldehyde (set III). Each of the triplicate subsamples was split into three treatments and either flash frozen in liquid nitrogen [N2(l)] and stored at 86°C (subsample A) or 20°C (subsample B) or stored directly at 20°C (subsample C). The remaining samples from sets II and III were stored at 4°C. Slides were prepared for enumeration of viral abundance at time zero (t=0) and 7 days (t=7). Glut, glutaraldehyde; Form, formaldehyde.
|
|
|
|---|
Effect of aldehyde fixation and storage at 4°C.
One of the major benefits reported for the use of SYBR Green I was its compatibility with fixatives (16). However, in both 16-day and 48-h experiments viral abundance decreased dramatically and rapidly in glutaraldehyde- or formaldehyde-fixed natural water samples stored at 4°C (Fig. 2). By 1 h postfixation, the abundance had decreased an average of 16.4% ± 5.2% (n = 6), and by 4 h, the viral abundance was 20 to 35% lower than that at zero time (Fig. 2B). Different concentrations of fixative did not result in significant differences in decay rates (P < 0.05, as determined by the slopes test; n = 41) (data not shown). The average rates of decay for glutaraldehyde- and formaldehyde-fixed samples over the first 2 h were 0.12 and 0.13 h1, respectively. The rate of decay decreased markedly with time and was almost 1 order of magnitude less after 8 h than during the period immediately following fixation. By 16 days the average viral abundance had decreased by 72.2% (standard deviation, 6%; n = 6) (Fig. 2A). The amounts of the decrease were not significantly different between the fixatives for the 48-h and 16-day experiments (P = 0.163 and P = 0.865, respectively, as determined by the paired t test). These results indicate that aldehyde fixation of samples followed by storage at 4°C for even a few hours results in large underestimates of viral abundance by EFM. This may be one of the reasons that the Yo-Pro-1 method was found to yield estimates of viral abundance that were 1.2- to 7.1-fold (average, 3.0-fold; n = 12) higher than estimates based on transmission electron microscopy (12), whereas the estimates obtained by using SYBR Green I averaged only 1.28-fold higher than transmission electron microscopy-based counts (16). It is possible that microwave radiation may reduce the decay (26), although we did not test this in our study; moreover, Xenopoulos and Bird caution that even when microwave radiation is used, the samples must be processed within a few days.
![]() View larger version (16K): [in a new window] |
FIG. 2. Percentages of the initial virus abundance for 0.5% glutaraldehyde-fixed ( ) and 2.0% formaldehyde-fixed ( ) samples collected from Jericho Beach Pier and stored at 4°C. Subsamples were collected over time, and the viruses were enumerated by using SYBR Green I staining. The percentages of the initial virus abundance estimates are shown over time for slides prepared immediately and frozen at 20°C for 0.5% glutaraldehyde-fixed ( ) and 2.0% formaldehyde-fixed () samples. Models of decay for glutaraldehyde- and formaldehyde-fixed samples are indicated by solid and dashed lines, respectively (see text for equations). (A) Long-term experiment conducted over 16 days. (B) Short-term experiment conducted over 48 h. The error bars indicate standard deviations for triplicate samples.
|
In our experiments the loss of viruses over time was not well described by a single rate constant. For example, the rate of loss of viruses in glutaraldehyde- and formaldehyde-fixed samples decreased from 0.13 and 0.17 h1 during the first hour to ca. 0.01 h1 between 24 and 48 h. This suggests either that a combination of mechanisms contributed to the decrease in viral abundance in fixed samples or that specific viruses decay at different rates.
Correcting for decay in fixed samples.
Although the rates of loss of the viruses changed with storage time, the rates were similar for viral communities collected from different environments. For example, 7 days after fixation ca. 40 to 60% of the viruses remained in refrigerated samples from the Jericho Beach Pier (Fig. 2A), the Fraser River (Fig. 3B), and Burrard Inlet (Fig. 3C). Given that the viral communities were likely different in the different habitats, this finding suggests that the rate constants for decay in aldehyde-fixed samples, although variable over time, might be quite robust among samples. This is consistent with previous findings (22) which showed that decreases in bacterial cell numbers in glutaraldehyde-preserved seawater samples are predictable.
![]() View larger version (20K): [in a new window] |
FIG. 3. Virus abundance estimates for slides prepared at time zero (t=0) from unfixed, 0.5% glutaraldehyde-fixed, and 0.74% formaldehyde-fixed samples. Estimates of abundance are also shown for day 7 (t=7) for samples fixed for 30 min or left unfixed and stored at 20°C or flash frozen in liquid nitrogen and stored at 20 or 86°C. Fixed samples were also stored at 4°C. Light grey bars, SYBR Green I-stained slides; dark gray bars, SYBR Gold-stained slides; black bars, Yo-Pro-1-stained slides. Where data are not shown, the treatment was not performed with the sample. The error bars indicate standard deviations for triplicate subsamples. (A) Jericho Beach Pier; (B) Fraser River; (C) Burrard Inlet. Glut, glutaraldehyde; Form, formaldehyde; Liq N, liquid nitrogen.
|
There are several plausible explanations for the decreases in viral abundance in stored fixed samples. Proteases may break down viral proteins, as has been suggested for bacteria in glutaraldehyde-fixed samples (10). Alternatively, aldehydes may cross-link the viral proteins and interfere with stain penetration. Finally, viruses may attach to each other, bacterial cells, or other particulates, and this may reduce the number of free viruses. Regardless of the mechanism, the viruses can no longer be distinguished as individual particles by EFM or flow cytometry (5, 13).
The high rate of loss of viruses in aldehyde-fixed samples is likely the reason that we obtained variable results and other workers have reported lower estimates of viral abundance when the protocol for SYBR Green I staining (16) was used. For example, Bettarel et al. (2) indicated that the estimates of viral abundance made by using the original Yo-Pro-1 method (12) were invariably and significantly higher than those obtained by using SYBR Green I. However, their slides were made from samples fixed in 0.74% formaldehyde and stored for up to 1 week.
Processing of unfixed samples and storage of slides.
We found no significant difference (P > 0.01) between estimates of viral abundance made with fixed and estimates of viral abundance made with unfixed samples, provided that the slides were made immediately. However, it was not possible to use SYBR Green I to stain all unfixed samples (e.g., Burrard Inlet samples) due to occasional sample-dependent high background fluorescence. Counts from time zero for natural samples from the three locations analyzed by using analysis of variance showed no significant difference among stains (P > 0.01), demonstrating that accurate estimates of viral abundance can be made with aldehyde-fixed samples, if the slides are made immediately.
There was no decline in viral abundance during the 16-day experiment on Yo-Pro-1-or SYBR Green I-stained slides stored at 20°C. This finding agrees with studies showing that Yo-Pro-1-stained slides can be stored at 20°C for at least 4 months (12) and with our unpublished data showing that accurate estimates of viral abundance can be made with Yo-Pro-1-stained slides stored for 1 year. Hence, slides do not need to be processed immediately after preparation.
Alternative preservation methods.
As storage of aldehyde-fixed samples at 4°C resulted in rapid decay of viral particles, we explored freezing at 20°C, flash freezing in liquid nitrogen and storage at 20°C, and flash freezing and storage at 86°C as alternative methods to preserve unfixed and fixed samples. Of the three approaches, only flash freezing followed by storage at 86°C produced acceptable results. This is also the protocol that has been recommended for storage of samples for viral enumeration by flow cytometry (6, 13). This method of storage has been found to provide reliable estimates of abundance by flow cytometry for up to 6 months of storage (6). Viral abundance decreased in samples that were flash frozen and stored at 20°C or were stored directly at 20°C (Fig. 3). An exception was unfixed samples from Burrard Inlet that were flash frozen and stored at 20°C, for which the estimates of viral abundance were about double those at time zero (Fig. 3C). The reason for this difference is unknown. Estimates of viral abundance determined by using SYBR Green I or SYBR Gold staining did not decrease for unfixed or 0.5% glutaraldehyde-fixed samples that were flash frozen in liquid nitrogen and stored at 86°C. The viral abundance for samples from Jericho Beach and the Fraser River that were fixed in 0.74% formaldehyde, flash frozen, and stored at 86°C was significantly different from the viral abundance for unfixed and 0.5% glutaraldehyde-fixed samples that were stored in the same way (P < 0.01). This finding is similar to the findings of experiments performed with viral isolates in which the flow cytometric counts decreased for formaldehyde-fixed samples (5). Flash-frozen unfixed samples showed lower abundance when they were stained with Yo-Pro-1 than when they were stained with SYBR Green I or SYBR Gold.
Recommendations.
Whenever possible, slides should be prepared immediately with either unfixed or fixed samples and frozen at 20°C until they are processed, because viral abundance declines rapidly in fixed samples stored at 4°C. Slides made immediately with Yo-Pro-1, SYBR Green I, and SYBR Gold yielded estimates of viral abundance that were indistinguishable from each other. Nonetheless, the suitability of the staining procedures should be checked for the type of samples being analyzed because not all stains were suitable for all samples. If immediate slide preparation is impractical, flash freezing 0.5% glutaraldehyde-fixed or unfixed water samples in liquid nitrogen, followed by storage at 86°C, allows accurate estimates of viral abundance to be made by using SYBR Green I or SYBR Gold staining. Consequently, determinations of viral abundance made with fixed samples should be considered large underestimates unless stained slides were made immediately after fixation or the samples were flash frozen in liquid nitrogen and stored at 86°C. If the protocols outlined above are carefully followed, EFM provides practical and accurate estimates of viral abundance in natural samples.
This study was supported by a grant from the Natural Science and Engineering Research Council of Canada to C.A.S.
|
|
|---|