Applied and Environmental Microbiology, October 2005, p. 6086-6095, Vol. 71, No. 10
0099-2240/05/$08.00+0 doi:10.1128/AEM.71.10.6086-6095.2005
Lawrence Livermore National Laboratory, Livermore, California,1 National Institute for Occupational Safety and Health, Cincinnati, Ohio2
Received 2 March 2005/ Accepted 22 April 2005
|
|
|---|
|
|
|---|
Many tests to diagnose tuberculosis in patients are currently used; however, the actual techniques used depend on the setting and vary in their sensitivity, specificity, speed, and cost (12). Collection of samples for traditional culture methods on solid media are considered the standard and can range from simple and inexpensive (sputum swabs) to difficult and painful (gastric lavage). Unfortunately, M. tuberculosis is a slow-growing Mycobacterium for which growth-dependent tests are time-consuming and can take up to several weeks for reliable results. The time requirement of such assays dangerously lags the pace of life in modern urban societies. In addition, naturally occurring aerosolized environmental mycobacteria may not be viable, and those that are viable may not be culturable (22, 42, 47). On the other hand, the Ziehl-Neelsen diagnostic method is one of the simplest and most rapid of techniques and uses staining and smear microscopy to detect acid-fast bacilli. However, it is insensitive, requiring 104 bacilli/ml of liquid, and is characterized by low specificity since it cannot differentiate M. tuberculosis from other mycobacteria. Serological techniques such as antibody-based enzyme immunoassays and molecular amplification methods such as PCR are much more specific, are highly sensitive, and can be performed more rapidly but still take hours to perform (17, 45).
Apart from genetic molecular information, some important chemical components have been used to discriminate between M. tuberculosis and other mycobacteria and/or microorganisms. Identification of tuberculostearic acid by gas chromatography-mass spectrometry has been shown to be a good marker for diagnosis of M. tuberculosis; however, this lipid can also be found in M. bovis, M kansasii, and M. africanum (24). The detection of the glycolipid, lipoarabinomannan, which composes up to
15% of the dry weight of the cell, has also been demonstrated for diagnostic purposes but again is not unique to M. tuberculosis (15). More generally, determination of the specific cell protein composition has been popular and shown to be useful for some discrimination of various species of microbes using matrix-assisted laser-desorption ionization mass spectrometry of intact microorganisms (3, 7, 23). More specifically, nuclear magnetic resonance spectroscopy (48), mass spectrometry (20), and high-performance liquid chromatography (4) have been used to obtain profiles of mycolic acids, which make up 40 to 60% of the dry weight of mycobacteria (2) and have been shown to be fairly specific to many species of mycobacteria.
Some of these methods can be used to help detect and identify airborne M. tuberculosis if coupled to appropriate air collection devices utilizing suitable substrates. Aerosolized M. bovis BCG and the avirulent M. tuberculosis H37Ra, surrogates for pathogenic M. tuberculosis, have previously been successfully sampled in a laboratory bioaerosol chamber (CDC-NIOSH, Cincinnati, OH) (35-37). These studies used viable Andersen six-stage samplers and eight-stage micro-orifice uniform deposit impactors containing Teflon filters for particle size fractionation. Disposable filter cassette samplers were also used as simple, sensitive filter collection samplers for airborne mycobacteria. The mycobacteria were readily detected by using PCR. Another study investigated aerosols of artificial saliva containing M. avium, M. intracellulare, and M. tuberculosis that were generated in a class III safety cabinet and sampled and analyzed by using viable Andersen six-stage samplers and viable all-glass impingers (21). Some of these techniques were subsequently applied, in the field, to the detection and characterization of mycobacteria associated with whirlpools at a large public facility (37). Also for the first time, the size range and concentrations of infectious tuberculosis aerosols were directly measured from culturable cough-generated aerosols of M. tuberculosis from infected human test subjects by using Andersen six-stage samplers with culturing (6). In that study, concentrations of airborne M. tuberculosis particles were measured at approximately 1 to 2 CFU/liter of air when sufficiently infected individuals repeatedly coughed into a small chamber for 5 min. All of these studies used methods that required only a few minutes for sampling the aerosols. Subsequent offline analysis for detection and characterization took approximately a few hours using the nonculturing techniques and up to a few weeks using the culturing techniques.
A rapid, reagentless on-line analytical technique to sample and detect bioaerosols, at the individual particle-level of resolution, called bioaerosol mass spectrometry (BAMS), is being developed at Lawrence Livermore National Laboratory for national security and public health applications that require detection of airborne bio-agents such as vegetative bacteria, bacterial spores, viruses, and biological toxins (8, 34, 41). Although the size and cost of BAMS is currently greater than some other techniques, it will be useful where real-time, unmanned, continuous operation is required, and development of a smaller and less-expensive version is under way. The BAMS technique is based on single-particle aerosol mass spectrometry (11, 18, 19, 25, 29, 30, 43, 44), which has previously been applied to the analysis of a range of biological particles (13, 16, 31, 38, 39). This is the first time it has been used to analyze M. tuberculosis. In BAMS, many individual airborne particles are sampled per second directly from the air and accelerated to size-dependent speeds in a vacuum system, where the individual particles are then individually sized by the time delay measured between two laser scattering events. The molecules in the particles are then desorbed and ionized using high-power 266-nm laser light, and the fragments are chemically analyzed by using dual polarity time-of-flight mass spectrometry (TOF-MS). BAMS has already been shown to be a promising technique for the detection of Bacillus spores of interest to biodefense applications, and the differentiation of two Bacillus spore species was demonstrated (8). BAMS is currently being developed further as an analytical method for the rapid, noninvasive diagnostics of infectious disease in respiratory effluent. Such a method would permit a more rapid response to public health concerns associated with diseases, such as tuberculosis and influenza.
Here we present the first results from an evaluation of BAMS for the rapid real-time detection of airborne mycobacteria using aerosolized forms of the avirulent variant M. tuberculosis H37Ra and M. smegmatis in a bioaerosol chamber. The signatures of these organisms are also compared to other gram-positive vegetative bacteria, B. atrophaeus and B. cereus. For the first time, the accuracy and sensitivity of BAMS is quantitatively established using an aerodynamic particle sizer (APS), viable Andersen six-stage samplers, and disposable filter cassette samplers, which permitted direct counts of the collected airborne mycobacteria cells.
|
|
|---|
B. atrophaeus (ATCC 9372) and B. cereus (ATCC 14579) were prepared as vegetative cells (i.e., nonspores). Stocks of these organisms were cultured in Nutrient Broth (Difco Laboratories, Detroit, MI) from a single colony isolated from a streak plate until turbid (after approximately 12 h). The cells were then subcultured at a 1:200 dilution and grown overnight in Nutrient Broth at 37°C in a shaker incubator. Finally, the cells were then centrifuged at 8,000 x g for 10 min and washed with sterile distilled water. This washing step was conducted three times.
APS.
An APS (model 3321; TSI, Inc., St. Paul, MN) was used to independently measure particle sizes and particle concentrations in the aerosol sampled in experiments characterizing the various stages of the BAMS system and in the aerosol sampled in the bioaerosol chamber experiments (both described below). The APS is a widely used technique that measures the concentration and aerodynamic diameter of individual particles (0.5 to 20 µm in diameter) by sampling and accelerating them to size-dependent speeds measured by two continuous wave lasers. For these experiments, the APS was run in the summed mode. Particles measured in bin 1 (<0.523 µm in diameter) were not used. It has been shown that the APS TSI 3221 particle counting efficiencies are roughly constant at
50% in the particle size range of this study compared to impactor sampling (32); therefore, measured APS concentrations were corrected by a factor of 2.
BAMS: overview.
An aerosol time-of-flight mass spectrometer (model 3800; TSI) based on a transportable design (11) was used with modifications, as previously described in detail (8, 34, 41) and referred to here as BAMS (Fig. 1A). Briefly, airborne particles and air were sampled into a two-stage virtual impactor (VI, MVA-400; MesoSystems, Kennewick, WA) at 400 liters/min. This was used to concentrate particles with diameters greater than 2 µm by a factor of 120 into the airflow that was sampled into the BAMS vacuum system. The VI stage was designed for an output minor flow of 3 liters/min; therefore, since only 0.9 liters/min of the minor flow was sampled into the BAMS vacuum system, the remaining flow was made up by a split off to a small vacuum pump. This made the effective sampling rate of BAMS approximately 120 liters/min. Within the BAMS apparatus, the preconcentrated aerosol was sampled into a differentially pumped vacuum system through a converging nozzle. Particles were accelerated into a high-vacuum region according to their aerodynamic diameters and individually sized by the time delay measured between two scattering lasers when a single particle passes through them. The measured particle speed was used to predict its arrival in the center of the ion source of a dual-polarity reflectron time-of-flight mass spectrometer. In the ion source, molecules from the particle were desorbed and ionized by using a frequency-quadrupled, Q-switched Nd:YAG laser. The laser power used was
660 µJ per pulse, and the laser beam diameter in the center of the ion source was
400 µm in diameter, resulting in a laser fluence of
0.5 J/cm2 (41). Both positive and negative ions were simultaneously extracted in opposite directions and analyzed by TOF-MS for single particle chemical analysis.
![]() View larger version (34K): [in a new window] |
FIG. 1. Schematic of aerosol generation, sampling, and detection experimental setups for BAMS mass spectral signature development using BAMS without its virtual impactor (VI) stage (A), BAMS' virtual impactor (VI) stage efficiency measurements (B), BAMS particle sampling and tracking efficiency measurements using BAMS without its VI stage (C), and the CDC-NIOSH bioaerosol chamber experiments (D).
|
BAMS: sampling efficiency measurements.
Particle sampling and detection efficiencies of the various stages of BAMS were determined to enable the calculation of absolute particle concentrations from BAMS measurements. Aerosol preconcentration efficiency by the VI stage was characterized by continuous aerosolization of monodisperse polystyrene latex microspheres (PSLs; Duke Scientific) into a small aerosol chamber (38 liters) using a collision nebulizer, as shown in Fig. 1B. A total chamber flow of
400 liters/min was set by the VI, which comprised of
3 liters of aerosolized PSLs/min and
397 liters of filtered room air/min. An APS was used to measure the aerosol concentrations. Here, only the internal nozzle of the APS was used to sample directly from the chamber at 1 liter/min, and the 4 liters of filtered sheath air for the APS/min was sampled from the room air. In order to calculate the overall preconcentration of particles with a specific aerodynamic diameter, aerosol concentrations measured in 1 liter/min of a total of 3 liters/min VI exit minor flow (Fig. 1B, solid line) was compared to the aerosol concentrations measured before entering the VI (dashed line).
The BAMS particle sampling and tracking efficiency for the stages after the VI was also characterized. Monodisperse PSLs of 0.5 to 5.0 µm in diameter were aerosolized into a small aerosol chamber (19 liters) using a collision nebulizer, as shown in Fig. 1C. BAMS and an APS simultaneously sampled from the chamber aerosol. The number of particles successfully tracked and sized by BAMS was compared to the particle concentration measured by the APS.
The ability to produce mass spectra from individual Mycobacterium particles was determined by measuring the fractional hit rate of the particles. The fractional hit rate was defined as the fraction of particles whose molecules were desorbed, ionized, and produced sufficient ions for a mass spectrum out of all of the particles that were successfully tracked and sized in BAMS. This rate is dependent on the particle type and size.
Particle transport efficiencies were also experimentally measured in the sample tubing (3/8-in.-inner-diameter copper; Swagelok; Oakland Valve, Oakland, CA) that led from the bioaerosol chamber to the APS and that led from the virtual impactor to the converging nozzle of the BAMS system. The particle losses for both systems were determined to be <5%.
Bioaerosol experiments: general protocol.
A typical bioaerosol chamber experiment was conducted as follows. Details are provided in the sections provided below. A Mycobacterium strain was cultured and washed. The concentration of mycobacteria in the suspension was determined. The suspension was diluted in sterile water to a desired concentration and transferred to a Collison nebulizer. Aerosolization began when clean air flow through the Collison nebulizer was started and was continued for 20 min. Sampling by all of the techniques was started when aerosolization began. The viable Andersen six-stage sampler was run for only the first 60 s of aerosolization. The BAMS, APS, and filter cassette samplers were run for the full 20 min of aerosolization.
Bioaerosol experiments: bioaerosol chamber.
The overall experimental setup at NIOSH (Cincinnati, OH) is depicted in Fig. 1D. As previously described (35-37), the NIOSH bioaerosol chamber (Torrington Research Company, Torrington, CT) was constructed of stainless steel and measured 30 in. by 30 in. by 19 feet (3,363 liters). The biochamber contained baffles downstream of the aerosol inlet in order to generate a uniform bioaerosol concentration within the chamber. The supply and exhaust air were purified by passing the air through prefilters and high-efficiency particulate air filters. The velocity throughout the cross-sectional area of the sampler location was uniform and determined to be approximately 100 feet/min (21,238 liters/min).
Bioaerosol experiments: microbial aerosol generation.
For both M. tuberculosis H37Ra and M. smegmatis, aliquots of the washed cells were diluted to three final concentrations of
1 x 107,
5 x 107, and
1 x 108 cells/ml. Approximately 50 to 100 ml of microbial suspension was placed in a six-jet modified MRE-type Collison nebulizer (BGI, Inc., Waltham, MA). Filtered compressed air (138 kPa) was used to generate the bioaerosols for 20 min. Negative controls for all samplers were obtained by operating the chamber under conditions similar to the microbial aerosolization experiments, omitting the microbial suspension.
Bioaerosol experiments: Andersen six-stage sampling and analysis.
A viable Andersen six-stage particle sizing sampler (Graseby Andersen Division, Smyrna, GA) positioned directly inside the bioaerosol chamber (Fig. 1D) was used to quantify the number of culturable airborne mycobacteria and for the determination of the aerodynamic diameter 50% cut point range of the aerosolized mycobacteria. The device sampled at a flow rate of 28.3 liters/min and was loaded with Bacto Dubos oleic agar plates (Difco). Sampling was conducted for 1 min during the initial phase of the mycobacteria bioaerosol generation period. The sampling time was limited to 1 min because longer sampling times at the cell aerosol concentrations generated would overload the sampler, preventing the ability to quantify. After sample collection, the agar plates were removed and incubated at 37°C in 7.5% CO2 for 2 to 3 weeks. The plates were examined daily for the presence of mycobacterium colonies. Visible colonies were manually counted on a model C-110W colony counter (New Brunswick Scientific Co., Edison, NJ). A positive hole conversion table was applied to correct for statistical effects (1).
Bioaerosol experiments: filter cassette sampling and analysis.
A three-piece filter cassette sampler (Gelman Sciences, Ann Arbor, MI) positioned directly inside the bioaerosol chamber (Fig. 1D) was used for direct cell counts of airborne mycobacteria in the bioaerosol chamber. The cassettes contained 37-mm PTFE filters (Costar no. 130810) with a 1.0-µm pore size. The airflow rate through the filter cassette sampler was 3.7 liters/min, and sampling was conducted for 20 min. After sample collection, the filters were treated with SYBR Green I nucleic acid gel stain (FMC BioProducts), and the cells were directly counted using an Olympus fluorescence microscope (B&B Microscopes, Ltd., Warrendale, PA). SYBR Green I is a highly sensitive fluorescent stain that binds to nucleic acids in the bacterial cells.
Bioaerosol experiments: BAMS sampling.
BAMS measurements of bioaerosol chamber samples were sampled directly, with the VI stage situated directly inside the chamber and positioned parallel to the aerosol flow path as depicted in Fig. 1D. The system sampled at a flow rate of 400 liters/min for 20 min. The first minute of BAMS sampling was used for comparison to the Andersen six-stage sampler and APS measurements. The full 20 min of BAMS sampling was used for comparison to the filter cassette sampler measurements.
Bioaerosol experiments: APS sampling.
APS measurements of bioaerosol chamber samples were made by using a 3/8 in.-inner-diameter copper sampling tube (Swagelok), which was situated directly inside the chamber and positioned parallel to the aerosol flow path as depicted in Fig. 1D. The APS sampled at a flow rate of 5 liters/min for 20 min. The first minute of APS sampling was compared to the BAMS results.
Bioaerosol experiments: calculation of aerosol concentrations measured by BAMS.
Particle sampling and detection efficiencies at numerous stages of BAMS were quantitatively determined as described above. The preconcentration of aerosol particles by the VI stage of BAMS was characterized. The VI stage was designed to concentrate particles with diameters of greater than 2 µm in air by a factor of 120. Figure 2A shows the fractional efficiency with which the VI stage concentrated particles with diameters in the range of 0.5 to 5.0 µm. In addition, Fig. 2B shows the fractional efficiency by which particles are successfully tracked and sized in BAMS, as a function of particle diameter. Furthermore, Fig. 2C shows the fractional hit rate measured for a combination of 1,000 M. tuberculosis H37Ra and 1,000 M. smegmatis particles, as a function of particle diameter in the range of 0.5 to 2.0 µm. The fractional hit rate was defined as the fraction of particles whose molecules were desorbed, ionized, and produced sufficient ions for a mass spectrum out of all of the particles that were successfully tracked and sized. Using these efficiencies, the aerosol concentration was calculated as measured by BAMS in units of particles/liter of air by using the following equations.
![]() View larger version (18K): [in a new window] |
FIG. 2. Experimentally measured particle preconcentration, sampling, and detection efficiencies for the various stages of the BAMS. (A) Fractional efficiency with which particles are concentrated in the virtual impactor (VI) stage exit minor flow as a function of particle diameter [VIeff(dp)]. This is based on a theoretical concentration factor of 120. (B) Fraction of particles successfully tracked and sized out of all particles that are sampled by the BAMS after the VI stage as a function of particle diameter [TKeff(dp)]. (C) Fractional hit rate, defined as the fraction of particles whose molecules are desorbed, ionized, and produced a mass spectrum out of all particles successfully tracked and sized by the BAMS as a function of particle diameter [Heff(dp)]. This was measured with 1,000 M. tuberculosis H37Ra- and 1,000 M. smegmatis-containing particles.
|
![]() | (1) |
The concentration of particles classified as "vegetative cell" or M. tuberculosis H37Ra-containing particles was calculated as follows:
![]() | (2) |
Using these equations, the concentration of particles present in the bioaerosol chamber samples was calculated from the number of particles measured by BAMS.
Technique matrix.
An overview of the capabilities of the techniques used here is presented in Table 1. For instance, the APS was used to measure particle concentrations in air and their relative sizes (0.5 to 20 µm) but could not identify them. The SYBR green I fluorescent cell stain counts of filter cassette-sampled particles were used to directly count the viable and nonviable cells in the sampled air. The viable Andersen six-stage sampler was used to quantify the concentration of viable cells that are culturable and measure their aerodynamic size. BAMS was used to quantify total airborne particles and their sizes and identify particles containing viable and nonviable cells. Among these techniques, only the APS and BAMS performed real-time measurements on individual airborne particles.
|
View this table: [in a new window] |
TABLE 1. Overview of the capabilities of the techniques used in this study
|
|
|
|---|
1,000 single-particle BAMS mass spectra from
1,000 single airborne M. tuberculosis H37Ra, M. smegmatis, B. atrophaeus, and B. cereus particles. Averages are shown for clarity since laser shot-to-shot variations of individual mass spectral signatures were observed. An example of these fluctuations is shown by a series of individual particle mass spectra of M. tuberculosis H37Ra containing particles in Fig. 4. These variations may be due to fluctuations in absorbed laser energies and biological composition, as previously described for BAMS analysis of Bacillus spores (41). The overall single particle BAMS mass spectra of M tuberculosis H37Ra, M. smegmatis, B. atrophaeus, and B. cereus looked similar to each other (Fig. 3) A novel pattern recognition algorithm (8), which is similar to an ART-2a clustering algorithm previously described (33, 40), was used in the analysis of individual BAMS mass spectra and classified single particle types in real time. Used on its own, this data analysis scheme could not efficiently differentiate between these vegetative bacterial cells. Nevertheless, this scheme could be used to efficiently differentiate these organisms from other particle types, such as bacterial spores and environmental particles, and classify them as "vegetative cells." Once a particle was classified as a vegetative cell using the pattern recognition algorithm, a rule-based method using the presence of unique marker m/z peaks over a certain signal threshold was used to further classify the vegetative bacterial cell; this method was also previously described in detail (8). With the BAMS mass spectral signatures developed in the present study, it was found that BAMS with pattern recognition analysis classified M. tuberculosis H37Ra and M. smegmatis containing particles as vegetative cells at a rate of
93%. The remaining 7% of the spectra in the data set were either not able to be classified by the algorithm or were not spectra of actual vegetative cells and were classified as unknown. Using a unique marker peak at m/z 421 in the BAMS mass spectra of individual M. tuberculosis H37Ra-containing particles (Fig. 3), BAMS with rule-based differentiation further classified M. tuberculosis H37Ra at a rate of
66% at an absolute specificity compared to M. smegmatis. This means that when 1,000 individual spectra of M. tuberculosis H37Ra were compared to 1,000 individual spectra of M. smegmatis, the signal level of m/z 421 was above a threshold that positively identified 660 individual spectra as M. tuberculosis H37Ra, and none of the M. smegmatis spectra were misidentified as M. tuberculosis H37Ra. In addition, when similarly compared to the other gram-positive vegetative bacteria, B. atrophaeus and B. cereus, M. tuberculosis H37Ra was positively classified at rates of
73 and
69%, respectively, indicating the m/z 421 peak is potentially unique to M. tuberculosis H37Ra. This is the first time a potentially unique biomarker was measured in M. tuberculosis H37Ra at the single-cell and single-particle level in real time.
![]() View larger version (35K): [in a new window] |
FIG. 3. Average of 1,000 single particle BAMS mass spectra of airborne M. tuberculosis H37Ra (A)-, M. smegmatis (B)-, B. atrophaeus (C)-, and B. cereus (D)-containing particles. Many of the individual spectra of M. tuberculosis H37Ra-containing particles ( 66%) can be differentiated from the three other organisms based on the presence of a m/z 421 peak.
|
![]() View larger version (40K): [in a new window] |
FIG. 4. Series of single-particle mass spectra from M. tuberculosis H37Ra-containing particles randomly selected to depict the typical wide variation in the m/z 421 signal intensity. It should also be noted that a subset of single-particle mass spectra from M. tuberculosis H37Ra-containing particles ( 34%) contain no significant signal at m/z 421. Within panel A is the structure of deprotonated trehalose-2-sulfate, the tentative identification of the m/z 421 peak.
|
,
'-D-trehalose-2'-sulfate known as sulfolipid 1 (SL-1) (26, 28, 49), has previously been positively identified as a component of the virulent M. tuberculosis H37Rv and M. smegmatis cells (27). This disaccharide was detected in its deprotonated form at m/z 421 in negative ion mass spectra utilizing Fourier transform-ion cyclotron resonance mass spectrometry of crude bulk cell extracts, but its relative abundance was not determined (27). We suggest that the m/z 421 peak associated with M. tuberculosis H37Ra in the BAMS mass spectra is tentatively due to the deprotonated disaccharide, trehalose-2-sulfate ion (Fig. 4A). Two fragment ions, m/z 241 and 259, representing the fragmentation of the disaccharide at either side of the oxygen link, were also present. We believe the absence of m/z 421 in the BAMS mass spectrum of M. smegmatis could be due to a significantly lower abundance of trehalose-2-sulfate, one that is below the intracellular detection limit of BAMS, since M. smegmatis is known not to produce SL-1. In future work, synthesis of a standard of trehalose-2-sulfate will be pursued in order to help authenticate the identity of this m/z 421 marker peak and further explore the intracellular concentrations and the robustness of this biomarker and other potential biomarkers.
Quantitative analysis of bioaerosol chamber samples.
Figure 5A and B show an example size distribution of the raw number of particles measured and the resulting calculated concentrations for each particle size, regardless of their identity. These were measurements made by APS and BAMS, respectively, in the bioaerosol chamber sample generated in the first 60 s of aerosolization of
108 M. smegmatis particles/ml of suspension. It can be seen that the raw particle counts by BAMS significantly undercounted particles less than 1 µm in diameter. However, when equation 1 was applied to account for known BAMS sampling biases, this was partially corrected for in the final calculation of particle concentration over most of the detected particle size range. Table 2 lists the total particle concentrations in the bioaerosol chamber, regardless of identity, measured by the APS and BAMS over the first 60 s of the aerosolization of the three different concentrations of M. smegmatis suspensions. These values included all of the particles (>0.523 µm) generated during the experiments, with corrections for sampling biases in both systems. The particle concentrations measured by both APS and BAMS agreed fairly well, except for the range of particles with diameters of <0.7 µm. BAMS was relatively insensitive to particles in this smaller particle size range due to their low transmission efficiencies, as depicted in Fig. 2A and B. The results presented in Table 2 and the size distributions shown in Fig. 5A and B show that as a particle counter and sizer, overall BAMS remained fairly consistent with the APS.
![]() View larger version (23K): [in a new window] |
FIG. 5. Raw particle counts and calculated particle concentrations, regardless of their identity, as a function of particle diameter during the aerosolization of the high-concentration suspension (108 M. smegmatis particles/ml) into the bioaerosol chamber measured by the APS (A) and BAMS (B) in the first 60 s of aerosolization.
|
|
View this table: [in a new window] |
TABLE 2. Total particle concentrations measured in the bioaerosol chamber during the first 60 s of aerosolization of M. smegmatis suspensions
|
|
View this table: [in a new window] |
TABLE 3. Concentrations of vegetative-cell-containing particles measured in the bioaerosol chamber during the first 60 s of aerosolization of M. smegmatis suspensionsa
|
|
View this table: [in a new window] |
TABLE 4. Concentrations of M. tuberculosis H37Ra-containing particles measured in the bioaerosol chamber during the first 60 s of aerosolization of M. tuberculosis H37Ra suspensionsa
|
2 x 108 M. tuberculosis H37Ra particles/ml of suspension. The Andersen six-stage sampler measured an abundance of CFU measured at the 7.0-µm cutoff. Contribution of cells to this stage of the sampler may be attributed to clumping. BAMS did not measure any M. tuberculosis H37Ra particles in the size range greater than 3 µm or less than 1 µm in diameter. BAMS was relatively insensitive to particles in these larger and smaller particle size ranges due to their low transmission efficiencies, as depicted in Fig. 2B. |
View this table: [in a new window] |
TABLE 5. Concentrations of vegetative-cell-containing particles measured in the bioaerosol chamber during the full 20 min of aerosolization of M. smegmatis suspensionsa
|
|
View this table: [in a new window] |
TABLE 6. Concentrations of M. tuberculosis H37Ra-containing particles measured in the bioaerosol chamber during the full 20 min of aerosolization of M. tuberculosis H37Ra suspensionsa
|
![]() View larger version (29K): [in a new window] |
FIG. 6. Raw M. tuberculosis H37Ra CFU and particle counts and calculated M. tuberculosis H37Ra CFU and particle concentrations as a function of particle diameter during the aerosolization of the high-concentration suspension (2 x 108 M. tuberculosis H37Ra particles/ml) into the bioaerosol chamber as measured by BAMS (A) and the Andersen six-stage sampler (B) in the first 60 s of aerosolization.
|
0.7 to 2.0 µm in diameter, where the correction for the BAMS sampling bias is very sensitive to particle diameter and, therefore, is very susceptible to uncertainty. This is reflected in Fig. 5B. In order to improve the confidence of the BAMS measurements at these lower agent concentrations, particle counts within the system need to be increased and sampling biases across the size range of interest (0.8 to 10 µm) need to be reduced. A novel sampling inlet with better particle transmission and focusing over a wide particle size range is currently under development in order to address this problem.
Conclusions.
We presented here an initial step in the development of BAMS toward a rapid, stand-alone, airborne M. tuberculosis particle detector. This is the first time a potentially unique biomarker was measured in M. tuberculosis H37Ra at the single-cell and single-particle level in real time. A biological particle detector that could detect and identify various airborne bacterial spores (8, 34, 41), vegetative cells, viruses, and biological toxins is the overall goal of these studies. Below, we present some of the improvements we are currently developing in order for BAMS to mature towards a detector that is useful in more realistic environmental backgrounds expected in a multitude of diagnostic scenarios, such as dirty air and respiratory effluent. The first step, which we demonstrated here, is the feasibility of BAMS for the selective detection and discrimination of relatively pure M. tuberculosis H37Ra from M. smegmatis in a fairly clean aerosol background using a potentially unique metabolic marker. From the quantitative results in the present study, it can be estimated that BAMS can detect M. tuberculosis H37Ra containing particles at concentrations >40 M. tuberculosis H37Ra CFU/liter of air in 1 min in a background-free environment and compares well with the standard of cell culture. However, the overall detection limit of these clean particles alone may be improved by increasing the particle sampling efficiency through the development of an enhanced particle inlet. Furthermore, low intracellular concentrations of potential biomarkers and/or agent mixed, clumped, or coated with some background material, such as dust or sputum, can also interfere with efficient detection. As a result, we are currently developing more targeted laser desorption/ionization schemes and a higher sensitivity, higher-mass-range TOF-MS in order to improve the robustness of unique biomarkers.
This study was performed under the auspices of the U.S. Department of Energy by the University of California, Lawrence Livermore National Laboratory, under contract W-7405-Eng-48. Funding was provided by the Lawrence Livermore National Laboratory through Laboratory Directed Research and Development grant ERD-002 and the Technical Support Working Group within the Department of Defense.
|
|
|---|