Applied and Environmental Microbiology, February 2006, p. 1380-1387, Vol. 72, No. 2
0099-2240/06/$08.00+0 doi:10.1128/AEM.72.2.1380-1387.2006
Copyright © 2006, American Society for Microbiology. All Rights Reserved.
M. Finney,3
M. R. Fulford,4
E. Frandsen,5
E. Østergaard,5
J. M. ten Cate,2
W. R. Moorer,2
A. Mavridou,6
J. J. Kamma,6
G. Mandilara,6
L. Stösser,7
S. Kneist,7
R. Araujo,8
N. Contreras,8
P. Goroncy-Bermes,9
D. O'Mullane,10
F. Burke,10
P. O'Reilly,10
G. Hourigan,10
M. O'Sullivan,10
R. Holman,12 and
J. T. Walker3*
Department of Medical Microbiology, Academic Medical Centre, Amsterdam, The Netherlands,1 Department of Cariology, Endodontology and Pedodontology, Academic Centre for Dentistry, Amsterdam, The Netherlands,2 Centre for Emergency Preparedness and Response, Health Protection Agency, Porton Down, Salisbury SP4 0JG, United Kingdom,3 Shepton Mallet, Somerset, United Kingdom,4 University of Aarhus, Aarhus, Denmark,5 Technological Educational Institute, Athens, Greece,6 University of Jena, Jena, Germany,7 University of Barcelona, Barcelona, Spain,8 Schülke and Mayr, Norderstedt, Germany,9 University College Cork-National University of Ireland, Cork, Ireland,10 Leeds Dental Institute, Leeds LS2 9LU, United Kingdom,11 Department of Clinical Epidemiology and Biostatistics, Academic Medical Centre, Amsterdam, The Netherlands,12
Received 22 July 2005/ Accepted 28 November 2005
|
|
|---|
|
|
|---|
200 CFU · ml1 (equivalent to that required for dialysis water) (1). Currently, dentists across the world have little or no evidence-based guidelines to control bacterial numbers in DUWS. Typically, patients in the EU visit general dental practices (GDPs) every 6 months, with over 20 million visits per year in 1998 in one large EU country alone (5). During almost every visit, the patient and the dental health care staff are exposed to the water from DUWS. These medical devices have the potential to harbor opportunistic or frank pathogens, and Legionella pneumophila, Mycobacterium spp., Pseudomonas aeruginosa, and Candida spp. have been recovered from DUWS (4, 7, 28, 34, 46, 56). Exposure of dental personnel to such pathogens can be inferred from the finding that dentists have significantly higher antibody titers to L. pneumophila than individuals in other, equivalent employment sectors (10, 11, 24, 35), and asthma may be associated with occupational exposure to endotoxin in aerosols from contaminated DUWS (29). In addition, P. aeruginosa isolated from a DUWS was found to be responsible for the hospitalization of two patients following a visit to a dental surgery (19). The presence of pathogens has further implications when one considers that failure of the three-in-one hand piece antiretraction valve (32, 37) could result in microorganisms being transferred among patients (cross-infection).
A wide range of disinfectant products are now being developed for use in DUWS, and these have been evaluated using a variety of approaches (39, 48, 49), although rarely have these products been compared in a general dental practice setting. The aim of this study was therefore to evaluate and compare the levels of control achievable by the application of different disinfectants to DUWS in GDPs.
|
|
|---|
Modification of DUWS.
In the case of mains water systems, dental units were modified to facilitate the addition of a disinfectant to the water used in the DUWS by fitting an externally mounted purge system (Micrylium Laboratories, Inc., Phoenix, AZ). This was carried out by a qualified dental unit installer following the manufacturer's instructions. The system consisted of a 700-ml bottle and two switches; one to toggle between municipal and bottle water supplies and the other to turn the air purge on or off. Using this system, the municipal water supply could be bypassed and the bottle used to add a disinfectant intermittently to the DUWS, e.g., overnight or during the weekend, after which time the municipal water supply could be reconnected for use during treatment. Alternatively, if a disinfectant was to be used continuously, it was added to the bottle supply, which was subsequently used throughout patient treatment sessions. Air purging of the waterlines, if recommended by the disinfectant's manufacturer, could also be carried out using this system.
Sampling of DUWS.
A test period consisted of 4 to 5 weeks of baseline value measurement followed by 6 to 8 weeks of disinfection. This period was always preceded by a nondisinfecting period of at least 14 days to allow biofilm formation (48) and washout of residual product when multiple products were tested serially in one DUWS. In order to measure DUWS contamination, samples (100 ml) were taken weekly for 4 to 5 weeks before and for 6 to 8 weeks after disinfection. Samples were collected at approximately mid-morning from the distal outlets of the three-in-one syringes of all DUWS. Where possible, a 1-centimeter length of the waterline tubing supplying the three-in-one syringe was cut out for biofilm analysis (48). One tubing sample was taken during pretreatment baseline measurement, and one was taken after the disinfection period was finished. All the staff performing the disinfecting and sampling procedures were instructed to use the same methodology.
Water samples were processed as described previously (45). Approximately 100 ml of water sample was passed through a sterile nozzle into a sterile water bottle containing 0.1 g sodium thiosulfate to remove any residual disinfectant (Abinghurst Ltd., Northampton, United Kingdom). Samples were returned to the laboratory in a cool box (4 to 8°C) within 3 h and then filtered through 100-ml capacity, 0.2-µm analytical test filter funnels (Techware, Poole, United Kingdom) in order to recover the waterborne microorganisms. The membrane was removed from the funnel with sterile forceps and placed in a screw cap sterile container (Elkay Products Inc., Shrewsbury, MA). Microorganisms were washed from the membrane by vortexing the container for 1 min in 10 ml of sterile phosphate-buffered saline (PBS).
TVCs and detection of Pseudomonas aeruginosa in water samples.
Total viable counts (TVCs) and counts of Pseudomonas aeruginosa (as a marker of the presence of opportunistic pathogens) were carried out as described previously (45), using decimal dilutions of the water samples (in sterile PBS) on (i) yeast extract agar for environmental microorganisms (3), incubated at 37°C for up to 3 days, and (ii) CFC supplemented with SR103 for Pseudomonas spp. (26), incubated at 37°C for up to 48 h. Colonies were counted on day 3, and the TVC was calculated as CFU · milliliter1. Colonies on the Pseudomonas agar were confirmed as P. aeruginosa according to color, ammonia production, and oxidase activity. TVCs were used as the definitive measure of total microbial contamination of the water passing through the DUWS. This was compared with the U.S. guideline for DUWS of
200 CFU · ml1 as recommended by the American Dental Association (1). All of the participants used the same microbiological methods (46). The limit of detection was 10 CFU · ml1.
Analysis of biofilm accumulation.
Biofouling was assessed by measuring the TVCs of defined areas of the inner surface of the tubing as previously described (49). The DUWS tubing was sectioned horizontally and rinsed twice in nonflowing sterile PBS to remove nonadhered cells, and the surface biofilm was removed by scraping it into 1 ml of sterile PBS with a sterile dental probe. The TVC was then determined as described above, and the limit of detection was 10 CFU · cm2.
Evaluation of disinfectant agents in GDP DUWS.
Disinfectants (Table 1) were selected from a previous laboratory model evaluation study according to their ability to kill microbial cells and remove biofilm from the inner surfaces of DUWS tubing (47, 48). Products were supplied and used either intermittently or continuously (see below) according to the manufacturers' instructions. Treatment was administered each evening prior to the closing of the surgery, and on weekends, for up to 8 weeks. Where necessary, the disinfectant was removed from the system by flushing the following morning before patient treatment began. The removal of disinfectant was confirmed either by observing the loss of product color from the outflowing water or from a negative potassium iodide detection test for hydrogen peroxide-based products.
|
View this table: [in a new window] |
TABLE 1. Summary of active agents, concentrations, type of application, and sources of the products evaluated
|
For intermittent disinfection of mains water-supplied systems, the unit was disconnected from the mains water supply and connected to the bottle with disinfectant by an externally mounted purge system (see "Modification of DUWS" above). The product was administered as described above, and at the beginning of the next working day the DUWS was reconnected to the mains supply for use in daily practice. The disinfectants applied intermittently were BioBlue, Sanosil, Sterilex Ultra, and Ster4Spray (Table 1).
Continuous application of disinfectants.
The DUWS water reservoir bottle was filled with disinfectant and connected to the unit, and the disinfectant was run through the system for 2 minutes. The dental unit was then ready for daily use, and the disinfecting agent was present in the outflowing water throughout the day. The disinfectants used for continuous use were Alpron, Oxygenal, and Dentosept (Table 1). The products were left in situ overnight.
Statistical analysis.
The results for pre- and postdisinfection for each product were compared using a Mann-Whitney test for two independent groups of samples. To test for statistical differences between all products, a Kruskall-Wallis test was used. Mann-Whitney tests were subsequently used to compare products with all other products separately (SPSS 12.0.1 for Windows; SPSS, Chicago, IL).
|
|
|---|
200 CFU · ml1, while during treatment, 83% of all measurements (n = 926) were below the ADA guideline. Baseline values from units in Germany were excluded from this analysis because these DUWS had an integrated disinfection system which provided an automatic continuous disinfection of the unit, making it impossible to determine nontreated baseline values. |
View this table: [in a new window] |
TABLE 2. Criteria used for evaluation of efficacies of the disinfection products in all countries
|
200 CFU · ml1 (
2.3 log CFU · ml1) in outflow water of the DUWS (Table 1).
(i) Alpron.
Alpron was tested in a continuous mode in 37 DUWS (Table 2). The use of Alpron resulted in a significant decrease in microbial counts to below 200 CFU · ml1 (P < 0.01) in comparison to the baseline values prior to treatment (Fig. 1). Eighty-seven percent of treated water samples achieved the ADA guideline, and the change in median TVC between the baseline and treated water samples was 2.10 log CFU · ml1 (Table 2).
![]() View larger version (26K): [in a new window] |
FIG. 1. Median water TVCs before ( ) and after ( ) the start of disinfection of DUWS for all products. Median values are shown as black lines in the boxes. The boxes represent the interquartile ranges, and the whiskers indicate the minimum and maximum values up to 1.5 times the IQR (made with SPSS 12.0.1 for Windows). , outlier value (between 1.5 and 3 times the IQR); *, extreme value (more than 3 times the IQR); , statistically significant change.
|
(iii) Dentosept.
Dentosept was tested in 11 dental units (Table 2) in a continuous mode. There was a significant difference (P < 0.01) between the baseline and treatment values (Fig. 1), with 91% of the treated water samples having a TVC of
200 CFU · ml1 and a change of 3.11 log CFU · ml1 between the baseline and treated water samples (Table 2). Extreme and outlier values that were measured (Fig. 1) indicate the occurrence of occasional high values during treatment. An outlier value was defined as being between 1.5 and 3 times the interquartile ranges (IQR), while extreme values were more than 3 times the IQR.
(iv) Oxygenal.
Oxygenal was tested in 15 DUWS (Table 2) in a continuous mode. A significant difference (P < 0.05) between pre- and posttreatment TVCs was found, and 91% of the treated water samples had a TVC of
200 CFU · ml1, with a change between the median baseline and treated water samples of 1.24 log CFU · ml1 (Table 2). One extreme value that was measured in The Netherlands (Fig. 1) was due to a problem with the application of the product to the DUWS, and the dentist stopped participating in the study after this high value was measured.
(v) Sanosil.
Sanosil was tested in intermittent mode in 30 dental units (Table 2). A significant decrease (P < 0.01) between results for baseline and treated water samples was found. The percentage of treated water samples that achieved the ADA guideline was 83%, and the change in median TVC between the baseline and test water samples was 1.57 log CFU · ml1 (Table 2). The broad range of the error bars (Fig. 1), indicating a variable effect of the product, was caused mainly by two units in Spain in which high values were found during treatment; these values were defined statistically as outliers in the longitudinal study (Fig. 2).
![]() View larger version (31K): [in a new window] |
FIG. 2. Longitudinal analysis of water TVCs during 8 weeks of disinfection of DUWS. Disinfection starts at week 1; week 0 represents the median baseline value. Median values are shown as black lines in the boxes. The boxes represent the interquartile ranges, and the whiskers indicate the minimum and maximum values up to 1.5 times the IQR (made with SPSS 12.0.1 for Windows). , outlier value (between 1.5 and 3 times the IQR); *, extreme value (more than 3 times the IQR).
|
(vii) Ster4Spray.
Ster4Spray was tested in intermittent mode in five units (Table 2). The median TVC was lower than the ADA guideline after treatment, but this difference was not statistically significant (Fig. 1). The distance between the error bars, representing the range of the values, indicates an inconsistent effect of the product (Fig. 1). Sixty percent of treated water samples were found to have TVCs of less than 200 CFU · ml1, with a change of 0.69 log CFU · ml1 between the median values of baseline and treated water samples (Table 2).
Longitudinal analysis of total viable counts during treatment.
A week-by-week comparison of products was performed to see if products differed in their patterns of disinfection, as it was hypothesized that some products might show a more rapid or sustained effect on TVC reduction than others (Fig. 2). Dentosept gave a marked and consistent reduction in median TVC values during the treatment period. The use of Alpron and, to a lesser extent, Oxygenal also caused a steady decline in median TVCs over time, whereas the other products did not have this effect (Fig. 2).
Dental units in which a product was tested for 6 weeks or more were included in a longitudinal analysis of the weekly TVCs to assess the presence of solitary high values in an otherwise low-value series. Samples treated with products in a series that yielded a median TVC of greater than 200 CFU · ml1 were excluded from this analysis. A solitary high TVC value flanked by low values in the weeks before and after such an occurrence was found for 26 of 111 samples (23%) (Table 2). Of these DUWS, seven units (6%) showed a second high value during the same testing period. The analysis showed that this phenomenon occurred with all products tested. Alpron was the product least prone to incidental high values, with 26 out of 29 units showing no evidence of high solitary TVC values. BioBlue, Sanosil, and Oxygenal all suffered from these high values in about 20% of dental units. Single high TVCs were present in almost half of the units that used Sterilex and Dentosept and in three of the five units that used Ster4Spray (Table 2).
Biofilm.
Biofilm samples from 28 DUWS in two countries (Greece and the United Kingdom) were evaluated before and after disinfection with the products Alpron, BioBlue, and Sanosil. Comparison of baseline and posttreatment values showed a significant decrease in median TVCs (P < 0.01) for all products tested. Baseline median TVCs were 1.45 log CFU · cm2 for Alpron and 1.04 log CFU · cm2 for both Bioblue and Sanosil. All products reduced the median TVCs after the treatment period to levels below the detection limit, corresponding to a median percent reduction of >99% for all three products (data not shown).
Detection of pseudomonads.
Pseudomonads were detected in Germany only. Three units were positive before the product evaluation period, and two of those units tested positive during treatment as well. Counts ranged from 5 to 125 CFU · ml1 (data not shown). All pseudomonads found were confirmed to be Pseudomonas aeruginosa.
Ease of use and adverse events.
A number of practical issues associated with the use of particular products were identified. These issues included dissolution of tubing and a lightweight nonpolypropylene plastic bottle on one unit (Sterilex Ultra), blocking of the tubing in the dental unit (Sterilex Ultra, three units; Ster4Spray, two units), and foaming and brownish discoloration of the dental unit water and staining of dental equipment surfaces in four units (Alpron). All Micrylium purge dispenser systems fitted to mains-fed DUWS that participated in this study had to be repaired at least once due to mechanical failures.
Assessment of overall efficacy of disinfectants.
A statistically significant difference in median TVC reduction (P < 0.01) was found between products used either continuously or intermittently (Table 1). The baseline median values for these two groups showed no significant difference, with 2.54 log CFU · ml1 for continuously used products and 2.49 log CFU · ml1 for intermittently used products. Treatment median values were 0.83 log CFU · ml1 for continuously used products and 1.25 log CFU · ml1 for intermittently used products.
There were significant differences between agents in terms of water TVCs (P value from Kruskal-Wallis-H). Further statistical tests showed that Dentosept was significantly more effective than all other agents (P = 0.0001). Oxygenal was significantly more effective than all agents except Dentosept (P = 0.001).
|
|
|---|
Various products have now been developed to reduce the number of bacteria delivered to the patient by decontamination of the water used in DUWS (12, 13, 17, 18, 20, 31, 36, 38, 40, 49), and although these have been tested in a variety of settings, no multiproduct comparison in GDP surgeries has been done until now. In this trial, seven antimicrobial products approved for use in dental units were tested in seven EU countries for their ability to reduce TVCs to meet the ADA guideline of 200 CFU · ml1 in the outflowing water (1) (Tables 1 and 2). Other criteria examined were the presence of occasional high TVC values in a series of low values, material incompatibility issues, the extent of the reduction of water TVCs during treatment compared to baseline TVCs in the same unit, and, where possible, the effect of disinfectants on biofilm TVCs.
Alpron.
Alpron has been tested in a number of studies. Smith et al. (39) found that Alpron reduced the TVC to a level similar to that in drinking water when used continuously over a period of 6 to 13 weeks. In the present study, during continuous use of Alpron, the bacterial counts of the outflowing water were reduced to below the ADA threshold of 200 CFU · ml1 in 87% of samples. Biofilm samples also showed a significant reduction in TVCs when measured before disinfection and after 8 weeks of disinfection with Alpron. This agrees with a previous finding that Alpron completely removes the biofilm in dental tubing (48). Others have shown that Alpron is also effective in reducing TVCs to acceptable levels (i.e., <200 CFU · ml1) when used intermittently (25).
BioBlue.
Kettering et al. (15) tested BioBlue (then named Bio2000) using a discontinuous regimen and concluded that it was not able to reduce the TVC to below 200 CFU · ml1 when used in combination with tap water. Although BioBlue achieved levels of less than 200 CFU · ml1 in 74% of samples in the present study, in some DUWS the measurements fluctuated, indicating an inability of the product to consistently maintain the TVC at below 200 CFU · ml1 between disinfection periods. A factor that might explain these findings could be the presence of weakened biofilm remnants after overnight treatment that slough off during the day. In this study, BioBlue reduced biofilm TVCs by >99%, although in a previous study in a laboratory model, BioBlue achieved only a 53% reduction of the tubing's inner surface biofilm coverage (48).
Dentosept.
Overall, Dentosept was very effective in reducing TVCs and maintaining the microbial load to levels well below 200 CFU · ml1 during this study, although occasional incidental high values were found. This product delivered the greatest reduction in TVCs, which resulted in the combined lowest number of samples which failed to meet the ADA guidelines during the treatment period. This is consistent with previous data which showed that daily application of Dentosept reduced TVCs in dental units to meet the ADA guideline (14).
Oxygenal.
Oxygenal has been evaluated previously in a laboratory DUWS biofilm model (48) and demonstrated a 99.2% reduction of biofilm coverage of dental tubing, while no TVC could be detected in the outflowing water after a single treatment. In the present study, Oxygenal achieved a statistically significant reduction in TVC, resulting in only a small percentage of samples that exceeded the ADA guidelines during the longitudinal study; the extent of the decrease was not very large, probably as a result of the low baseline values (Fig. 1).
Sanosil and Sterilex Ultra.
Tuttlebee et al. (43) compared Sanosil (5 to 8 weeks) and Sterilex Ultra (7 to 20 weeks) in GDPs using once-weekly disinfection of the DUWS, and both products reduced TVCs significantly to below the ADA guideline of
200 CFU · ml1. Using microscopy imaging techniques, a significant reduction in biofilm coverage has also been observed (43, 48). Similar results were reported for both products in laboratory models of DUWS (22, 48). In the present study, Sanosil and Sterilex Ultra significantly reduced or maintained TVCs below ADA guidelines; Sanosil also had a significant effect on biofilm TVC, as described previously (48). Unfortunately, the application of Sterilex Ultra was stopped in three DUWS due to material incompatibility issues. The problem of blocking of DUWS tubing when using Sterilex Ultra has been described before (43) and may be caused by sedimentation of excess Sterilex Ultra that has not dissolved settling onto the tubing's narrow inner surface. In a previous study, Sterilex Ultra initially reduced bacterial counts to levels below the ADA guideline but failed to maintain low TVCs during an 8-week follow-up period (16). This finding concurs with the relatively high prevalence of incidental high values that was found in the present study (Table 2), which may have been caused by regrowth and detachment of the biofilm.
Ster4Spray.
Ster4Spray was tested in only five units and showed variable TVCs. The product seemed prone to incidental high values, but further research with more DUWS should be carried out in order to obtain more conclusive data. A previous study which tested six units reported that Ster4Spray reduced the TVC to below 200 CFU · ml1 in 90% of samples collected over 10 days (23). However, the number of units tested was also small, and the testing period was too short to be able to conclusively assess the long-term efficacy of the product.
Low median baseline values and percent reduction.
The antimicrobial efficacy of a disinfectant should be assessed by both the reduction in TVC (log CFU · milliliter1) and the mean percent reduction in absolute TVC (Table 2). This is because even a relatively small reduction in log CFU · ml1 can correspond to a high percent reduction in absolute TVC when the initial (baseline) median TVCs are low. This effect was clearly visible in the Oxygenal data in Table 2, where a mean reduction in log CFU · milliliter1 of 1.24 corresponds with an absolute TVC reduction of 97%. Median baseline TVC values in both countries where Oxygenal was tested (Ireland and The Netherlands) were lower than 200 CFU · ml1, and in this case the percent reduction was a more accurate indicator of the efficacy of the product than the absolute reduction of medians in log CFU · milliliter1.
Longitudinal analysis of TVCs.
The incidental high TVC values found in the longitudinal analysis of the data during disinfection may have been caused either by contamination of the water sample by biofilm detaching from the inner surface of the dental tubing or by a regrowth of bacteria due to a lack of compliance by dental staff in applying the disinfectant according to the manufacturer's instructions. Compliance has been shown to be an issue when nonautomated procedures are in place (53). Several chair manufacturers, e.g., KaVo and Siemens, now integrate an automated water disinfection system with electronic monitoring in their DUWS. This reduces the occurrence of problems that may be the result of accidental noncompliance. The occurrence of the aforementioned occasional high TVC values has not been described before and can go unnoticed if a unit is tested only periodically. Such levels of bacteria might pose a threat to dentists and patients. They also indicate that the water quality of units should be tested at several time points so that an accurate assessment of the numbers of microorganisms in the outflowing water can be made.
The nature of this multicountry, multiproduct study using dental units of various design did not make it possible to have equal numbers of identical units using each disinfectant. However, the longitudinal analysis gave insights into the extent and consistency of the antimicrobial effects of particular agents. Continuous use of Dentosept (and to some extent Alpron and Oxygenal) produced a marked and sustained reduction in water TVC (Fig. 2).
Ease of use and adverse events.
A number of practical issues associated with the use of particular products were reported by the dentists. Alpron required a 3-hour time commitment for the three-stage initial application phase, and in some practices, its use resulted in foaming, staining, and a brown discoloration of the water. Ster4Spray and Sterilex Ultra caused blocking of the tubing in some DUWS. However, material compatibility issues are often recognized by manufacturers, who may subsequently recommend specific products to be used with their equipment; for example, the manufacturers of Sterilex Ultra advise against the use of nonpolypropylene bottles. Therefore, dental practitioners must consult with the manufacturer of their DUWS prior to introducing any chemical agent, as this may otherwise invalidate their warranty. Other studies have raised additional concerns. Brass coupling connectors used in domestic water systems can be corroded by too harsh a chemical treatment, and this may lead to leaks and failure of the DUWS (44). Disinfectants could come into contact with the oral cavity, and this might affect the adhesion of resins to both enamel and dentin, so that restorations could fail prematurely (36, 42). The risk of chemical exposure of the patient and health care workers also has to be considered when using disinfectant products in DUWS (17).
In GDPs supplied by the mains water supply, independent water dispensers (Micrylium, Toronto, Canada) were retrofitted to the mains supply line in order for disinfectants to be added to the DUWS. All dispensers had to be repaired or replaced at some time during the study. Bottle-fed units needed no adaptation, and few problems were reported.
Evaluation of product efficacy.
In conclusion, this study, in which a number of disinfectants have been compared under realistic conditions of use in GDPs in several countries, has identified products that reduced the TVCs in DUWS to levels recommended by the ADA. When further statistical analyses were performed, in which all agents were compared with each other, the most effective products were Dentosept and Oxygenal. However, Dentosept gave the most consistent and sustained antimicrobial effect over time. Overall, the continuously applied products performed better than the intermittently applied products.
We acknowledge the assistance of the disinfectant manufacturers for the supply of disinfectants and of the dental teams for their time and for allowing access to the DUWS.
No conflicts of interest are reported by any of the authors.