Applied and Environmental Microbiology, October 2008, p. 6437-6443, Vol. 74, No. 20
0099-2240/08/$08.00+0 doi:10.1128/AEM.01194-08
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

Defence Science and Technology Laboratory (Dstl), Porton Down, Salisbury, Wiltshire SP4 0JQ, United Kingdom,1 School of Biosciences, Geoffrey Pope Building, University of Exeter, Exeter, Devon EX4 4QU, United Kingdom2
Received 29 May 2008/ Accepted 12 August 2008
|
|
|---|
|
|
|---|
40% are greater than 7 µm, due to adherence to debris (27). Environmental microbial aerosols can arise from a number of sources, including food processing (13, 19, 40); water sources, e.g., contaminated air conditioning systems, shower heads, water faucets, and cooling towers (2, 42, 53); pesticide sprayers (1); and fungal-spore generation (15, 26). The importance of bioaerosols from a civilian biodefense perspective was highlighted by the U.S. anthrax mail attacks in 2001 (3). In health care settings, nosocomial transmission may occur by patients coughing or sneezing, generating respirable particles within the diameter range of 0.5 to 12 µm (6, 14); alternatively, pathogens can be reaerosolized after previous deposition on surfaces (46). Several important human and veterinary pathogens are transmitted by contaminated respiratory droplets (8, 46, 47). Evaporation of water from large respiratory particles results in the generation of smaller particles comprising the dried components, such as microbial biomass and mucin (37, 51). The rate of evaporation is dependent on the relative humidity (RH) and temperature of the air, air turbulence, and the constituents of the particle. The quantity of nonvolatile material within particles of >20 µm results in particles larger than 5 µm (37).
Particle size influences the site of deposition in the respiratory tract within humans and animals (21, 24, 29, 41, 45) and survival of the microbe in the aerosol (28, 50) and determines the length of time aerosolized microorganisms remain suspended within the atmosphere (25). Small particles around the size of a bacterium (1 to 5 µm) deposit in the alveoli, while larger particles (>10 µm) deposit in the upper respiratory tract (URT) (36, 41, 45). The importance of particle size to infection, microbial survival, and deposition is clear; however, the ability to assess bioaerosols with respect to particle size under defined experimental conditions is limited. Historically, the Collison nebulizer has been used to generate small-particle aerosols with a mass median aerodynamic diameter (MMAD) of 1 to 3 µm for animal exposure studies (34). Aerosols with particles greater than 10 µm have been produced by a spinning-top aerosol generator (STAG) connected to a drying column (9, 10, 11, 12, 21, 43). However, this method is difficult to carry out in microbiological containment facilities created for animal exposure studies. Alternative methods for generating large-particle aerosols would be beneficial.
Recently, a novel flow-focusing method of generating large-particle aerosols has been developed (17). Liquid is pumped through a capillary needle of defined diameter and drawn through a critical orifice. Upon application of an appropriate pressure from a compressed air supply, the focused liquid splits into particles of a defined size at a specific distance from the orifice of the aerosol generator. The orifice diameter is fixed; however, upon modulation of the liquid flow rate or gas pressure, particle sizes greater than 10 µm can be generated. The aerosols are produced with a geometric standard deviation (GSD) of 1.1 to 1.3, indicating particle uniformity (16, 17, 31). An important distinction between the two methods is the generation of wet and dry particles by the flow-focusing aerosol generator (FFAG) and the STAG, respectively. This technique may be applicable to the generation of large-particle aerosols containing microbiological material.
This study aimed to characterize 12-µm large-particle aerosols produced by the FFAG with respect to MMAD, particulate distribution within the particles (Escherichia coli cells, Bacillus atrophaeus spores, and fluospheres), bacterial survival, and deposition and clearance of the aerosols within the murine respiratory and gastrointestinal (GI) tracts. Comparisons with the Collison nebulizer, a device commonly used to produce small-particle microbiological aerosols for pneumonic models of disease in biodefense and health care research, are made.
|
|
|---|
Animals.
Female BALB/c mice (Charles River, United Kingdom) were used during the study and housed with free access to pelleted food and water within a rigid unplasticized polyvinylchloride (PVCu) fixed-film isolator at biological safety level 3 (BSL-3). The mice supplied for aerosol exposure studies were 6 to 8 weeks old and 20 to 25 g in weight. All mice were permitted to acclimatize for 1 week prior to any procedures. All procedures were performed in strict accordance with the Scientific Procedures (Animals) Act of 1986.
Aerosol generation and sampling.
The particulates comprised E. coli cells, B. atrophaeus spores (prepared as described above), or fluospheres (diameter of 1.1 ± 0.035 µm, volume of 0.697 µm3; Molecular Probes, Inc.) diluted to a concentration of 109 ml–1. Aerosol samples were collected for 1 min into an all-glass impinger (AGI-30; Ace Glass Inc., NJ) containing 10 ml of phosphate-buffered saline (PBS) at a flow rate of 12 liters min–1. The impinger samples were serially diluted and plated onto agar for enumeration as previously described (32). The small-particle aerosols were generated using a three-jet Collison nebulizer (13.5-µm-diameter jet holes) at a pressure of 26 lb/in2 (179,000 Pa) (34). For animal exposures, the Collison nebulizer spray was mixed with a secondary air supply of controlled RH of 45 to 50% within a Henderson apparatus (23) contained within a rigid PVCu half-suit isolator with a stainless steel base (Howorth Airtech Ltd., United Kingdom) at BSL-3. At 45 to 50% RH, predominantly 1-µm particles containing single microorganisms are generated (34). The total flow rate through the Henderson apparatus was 66 liters min–1. The large-particle aerosols were generated using an FFAG (75-µm-diameter orifice) purchased from Ingeniatrics Technologías (Seville, Spain). Tubing and connections were purchased from VWR International, United Kingdom. The FFAG relies on the formation of a stable microjet that disintegrates at a defined distance from a critical orifice (16, 17). For the generation of 12-µm particles, the system was operated at a pressure of 16 lb/in2 (110,000 Pa) derived from a filtered compressed air supply and the liquid supplied via a syringe pump (model 11 Plus advanced pump; Harvard Apparatus, United Kingdom) at a flow rate of 50 µl min–1. The generation of 12-µm particles for animal exposure studies was performed under operation at BSL-3. The parameters used while operating the FFAG are described above. In addition, air was mixed from a secondary air supply of controlled RH at 65 to 70% derived from the Henderson apparatus to prevent evaporation of the large particles. An elevated RH of 65 to 70% was used to ensure that evaporation of the 12-µm particles did not occur. The particle size distribution was determined by use of a Grimm no. 107 environmental dust monitor (Grimm Aerosol Technik GmbH & Co., Germany), which utilizes orthogonal laser light scattering at a 90° angle. The measurement range for particles was from 0.25 to >32 µm. Particle counts ranged from 1.0 x 106 to 2.0 x 106 liter–1. The aerosol was continuously sampled at a flow rate of 72 liter h–1, with the number and size of particles measured each minute. The aerosols generated are described by the MMAD, which is the particle diameter that halves the mass distribution; the GSD, which provides a representation of the uniformity of the particles and is obtained by dividing the MMAD by the particle size at either the 15.78% or the 84.13% probability; and the particle sizes below which 10% or 90% of the total particles produced in the aerosol occur.
Characterization of particulate distribution.
E. coli and B. atrophaeus spores were chosen as representative gram-negative vegetative cells and gram-positive spores, respectively. Fluospheres (diameter of 1.1 ± 0.035 µm; volume of 0.697 µm3) possess dimensions similar to those of bacteria and were used as surrogate bacterial particulates. Bacteria or spores were prepared and enumerated as described above before dilution with PBS to a final concentration of 109 ml–1. The bacterium or spore preparations were labeled with 1 mg ml–1 fluorescein isothiocyanate (FITC) (Sigma-Aldrich Ltd., United Kingdom) for 24 h at 4°C. Excess FITC was removed by multiple washes in distilled water with centrifugation steps (13,000 x g) until the supernatant was clear. The E. coli cells, spores, or yellow fluospheres were diluted with distilled water to a final concentration of 109 ml–1. The particulate suspensions (E. coli, spores, or fluospheres) were aerosolized and collected in a seven-stage Ultimate cascade impactor (35). A microscope slide thinly coated with 5% (vol/vol) gelatin was placed at each stage to collect the particles. The length of spray time was restricted to 5 s to prevent the coagulation of particles observed at higher spray times. The impacted particles leave an impression in the gelatin proportional to the original diameter of the particle. Only particles of a defined size range, from >1 to <32 µm, were collected on each of the seven stages. The slides were analyzed by use of an Olympus IX70 inverted laser confocal microscope (Olympus, Germany) under x60 water immersion. The fluorescent samples were excited at 488 nm and detected using either a 530-nm band-pass filter (FITC, fluorescein, green FluoSpheres) or a 565-nm long-pass filter (red FluoSpheres). Images were recorded and processed using a Fluoview imaging system (Olympus, Germany). On each slide, at least 200 particles were counted from 20 fields of view. The number of particulates contained within each particle was recorded. The diameter of the original particles prior to impaction into the gelatin was derived (in micrometers) from the scale provided by the imaging software. The actual size of the original particle was obtained by multiplication of the diameter of the impacted particle by a conversion factor of 0.59 (33). All experiments were performed in triplicate.
Determination of the effect of aerosolization on bacterial viability.
E. coli MRE162 was cultured to a density of 2.2 x 109 ± 0.13 x 109 CFU ml–1 in nutrient broth and split into equal 10-ml volumes for aerosolization using the Collison nebulizer or the FFAG for 1, 2, or 5 min. The aerosols were collected on fresh Whatman glass microfiber filters (grade GF/A, 47-mm diameter; Sigma-Aldrich Ltd., United Kingdom) by drawing compressed air through at a flow rate of 66 liters min–1. Immediately after collection, the filters were placed into 5 ml of PBS and vigorously shaken for 30 s to remove the deposited bacteria. Enumeration was done by serial dilution and spread plating of 100-µl aliquots in triplicate onto nutrient agar plates. The plates were incubated for 24 h at 37°C. The percent survival was calculated by dividing the number of bacteria obtained after aerosolization by the number of bacteria present preaerosolization. The resultant value was multiplied by 100 to derive the percentage.
Determination of site of deposition within the murine model.
Mice were nose-only exposed for 10 min to aerosols containing either E. coli MRE162 or FluoSpheres by use of the FFAG or the Collison nebulizer. The exposures were maintained at a constant temperature of 20 ± 0.5°C. The respiratory volume was obtained from Guyton's formula (20), and the retained dose was calculated via the method of Harper and Morton (21). The actual inhaled dose was derived from the nasal washes and dissected lungs, tracheae, esophagi, and stomachs from sacrificed mice. For each particulate, 10 mice were challenged with either large- or small-particle aerosols. Immediately after exposure, the mice were culled by irreversible anesthesia via intraperitoneal administration of 0.5 ml sodium pentobarbital. The respiratory tracts and digestive tracts were aseptically dissected. Nasal washes were collected by inserting a catheter (outer and inner diameters of 1.02 and 0.58 mm, respectively, and length of 40 mm; Harvard Apparatus Ltd., United Kingdom) into a small incision made in the trachea; the deposited materials in the nasal passages were washed out through the nostrils by use of 1 ml of PBS solutions containing final concentrations of 1 mM NaOH and 0.1% (vol/vol) Triton X-100. The lungs, trachea, esophagus, and stomach were removed and homogenized in PBS through a 0.1-µm cell strainer (VWR International, United Kingdom) by using the plunger from a disposable 2-ml syringe (VWR International, United Kingdom). Enumeration of bacterial counts was done by serial dilution and spread plating 100-µl aliquots in triplicate onto nutrient agar plates. The concentration of FluoSpheres within the homogenized tissue samples was determined by comparison against a standard curve of homogenized tissue samples spiked with known concentrations ranging from 102 to 108 FluoSpheres ml–1. Fluorescence was detected using a spectrofluorometer (excitation wavelength of 488 nm, emission wavelength of 520 nm; Molecular Devices, CA). Data were acquired and assimilated using SOFTmaxPRO software (Molecular Devices, CA).
Statistical analysis.
Aerosol distribution data are expressed as the standard errors of the means from either three or five replicate experiments. A paired t test was used for comparison between groups by use of the MiniTab Release 14 version 02 statistical package.
|
|
|---|
![]() View larger version (28K): [in a new window] |
FIG. 1. Representative mass distributions for the Collison nebulizer and the FFAG at the points of generation and inhalation. , Collison nebulizer, preinhalation (MMAD of 3.84 µm, GSD of 2.12); , Collison nebulizer, inhaled (MMAD of 2.51 µm, GSD of 1.28); , FFAG, preinhalation and inhaled (MMAD of 12.63 µm, GSD of 1.35).
|
|
View this table: [in a new window] |
TABLE 1. Effect of particulate type on the number incorporated per particle size fraction for the Collison nebulizer and the FFAGa
|
![]() View larger version (95K): [in a new window] |
FIG. 2. Representative confocal microscopic images of particles collected in the Ultimate cascade impactor from an aerosol (12-µm MMAD) generated by the FFAG containing FluoSpheres (a and d), B. atrophaeus spores (b and e), or Escherichia coli MRE162 cells (c and f). The imaged particles are 8 to 16 µm (a, b, and c) or 1 to 2 µm (d, e, and f). Particulates were aerosolized at a concentration of 109 ml–1. Arrows indicate empty particles.
|
![]() View larger version (67K): [in a new window] |
FIG. 3. Representative confocal microscopic images of (a) normally and (b) extremely loaded 12-µm particles collected in the Ultimate cascade impactor from an aerosol generated by the FFAG containing B. atrophaeus spores. Particulates were aerosolized in distilled water at a concentration of 109 ml–1.
|
![]() View larger version (10K): [in a new window] |
FIG. 4. Comparative survival of E. coli MRE162 cells after aerosolization. Black and white bars represent depositions of the Collison nebulizer and the FFAG, respectively. Averages were derived from three representative experiments.
|
![]() View larger version (12K): [in a new window] |
FIG. 5. Deposition in the murine model of aerosols generated by the Collison nebulizer and the FFAG containing (a) E. coli MRE162 and (b) FluoSpheres. Black and white bars represent deposition of the Collison nebulizer and the FFAG, respectively. Distribution data are as follows: for distilled water, MMAD of 12.36 µm and GSD of 1.38; for E. coli, MMAD of 12.11 µm and GSD of 1.33; and for FluoSpheres, MMAD of 12.36 µm and GSD of 1.39. Particulates were sprayed at a concentration of 109 ml–1. Error bars represent the standard errors of the means (n = 10).
|
|
|
|---|
An important consideration regarding deposition within the respiratory tract is the sizes of particles inhaled by the animals during exposure rather than present at the point of production by the device. Two profiles were obtained for the Collison nebulizer, corresponding to the distributions generated at the orifice of the device and at the point the animals inhaled the aerosol. Previous analysis of the particles produced at the orifice of the Collison nebulizer indicated a range of 0.25 to 25 µm (34). In this study, the mass distribution of the aerosol generated by the Collison nebulizer at the orifice was a close parallel. However, the aerosol that the mice inhaled was in the 1- to 5-µm range, due to evaporation within the Henderson apparatus at 45 to 50% RH. If evaporation of the 12-µm particles occurred during travel from the point of production to the animals' noses, then the particle distribution inhaled and hence the deposition profile would be altered significantly. Evaporation of the 12-µm-particle aerosol produced by the FFAG is prevented by feeding the aerosol directly into the exposure tube at 65 to 70% RH. Therefore, the aerosol to which the animals are exposed is identical to that generated at the orifice of the FFAG. Hence, within the exposure tube, the aerosol generated by the Collison nebulizer comprised a significantly greater number of smaller particles (1 to 5 µm), while the aerosol generated by the FFAG contained a greater number of larger particles (10 to 15 µm). The Collison nebulizer and the FFAG produced different mass distribution profiles, which enables the effect of particle size on deposition within the lower and upper respiratory tracts to be determined (Fig. 5).
The relationship of particulate number to particle size is important because it impacts the number of particles required to be inhaled to initiate infection (i.e., the lethal dose) or the quantity of material that can be deposited on surfaces. As 12-µm particles contain greater numbers of bacteria or spores than do 1-µm particles, it is evident that fewer particles need to be inhaled to reach the lethal dose and initiate infection. This is of course subject to the proviso that differential regional depositions within the respiratory tract due to particle size do not alter disease kinetics and severity. The volume of a spherical particle with a 12-µm diameter is 904.78 µm3. Assuming ellipsoidal morphology, the volumes of B. atrophaeus spores and E. coli cells have been determined as 0.273 ± 0.046 µm3 (5) and 0.33 µm3 (length of 1.4 to 1.6 µm; diameter of 0.65 ± 0.04 µm) (52), respectively. The number of unit spheres (particulates) that can be packaged into a larger volume (particle) is determined by sphere packing theory. Due to the rigidity of the spheres, the larger volume will not be filled because spaces occur between spheres, requiring consideration of fractional filling volume. Two packing arrangements permit the densest packaging of small spheres (particulates) into lattices within larger spheres (particles) with a fractional filling volume of 0.7405: face centered cubic and hexagonal close packing (7, 30). Hence, the maximum calculated numbers of E. coli cells, B. atrophaeus spores, or FluoSpheres packaged within a 12-µm particle are 2,030, 2,454, and 961, respectively. The actual particulate numbers observed were much lower (Table 1; Fig. 2), probably because a large portion of the matrix of the particle comprised the suspending medium from which the aerosol was derived. Due to aggregation, higher numbers of particulates were observed, particularly for B. atrophaeus spores (Fig. 3) but also occasionally for E. coli cells. Inhalation of these particles may have a significant effect on disease outcome and warrant consideration.
The survival of bacteria during the initial process of aerosolization in the apparatus will influence survival during extended aerosolization and thus infection upon deposition in the respiratory tract. Initially, mechanical and shear stresses are imparted by the actual act of aerosol generation within the device. Thereafter, bacteria are rapidly inactivated in situ due to a number of environmental stresses, including desiccation and UV radiation (27, 49, 50). In this study, temporally dependent significant differences between the numbers of culturable E. coli cells collected on filters after aerosolization by the Collison nebulizer and by the FFAG were observed (Fig. 4). It is currently unclear whether the increased survival observed with E. coli aerosolized by the FFAG was due to differential survival resulting from incorporation into larger particles or due to the reduced mechanical stress imparted during flow focusing. Interestingly, bacterial survival has been related to particle size. Bacteria survived better within 7-µm particles than within 1.1-µm particles, as a function of temperature, RH, and solar radiation (28, 50). This phenomenon is proposed to be due to larger particles containing higher numbers of bacteria in aggregates, enabling a fraction to survive deleterious stresses encountered during aerosolization. The outer layer of bacteria effectively is sacrificial, enabling the bacteria within the core of the aggregate to survive (4, 28); such aggregation of B. atrophaeus spores was observed in this study (Fig. 3). An interesting aspect of bacterial physiology not investigated in this study is the phenomenon of bacterial injury and the formation of "viable but nonculturable" cells during aerosolization (22, 44, 48). The physiological status of aerosolized bacteria may play an important role in aerosol survival, persistence, and, subsequently, infection with respiratory pathogens, and research to investigate this phenomenon with respect to particle size is under way.
The overall deposition trends for the aerosols generated by the Collison nebulizer and the FFAG are similar to previously published data in the murine model, although the CF1 strain was used rather than the BALB/c strain (41). For 1.09-µm particles, the fractions deposited in the lungs (pulmonary and bronchial), trachea, URT (skull and larynx), and GI tract were 26.4, 0.3, 8.6, and 64.7%, respectively. In this study, for the Collison nebulizer (1- to 3-µm particles), the fractions deposited in the lungs, trachea, URT, and GI tract were 8.04, 0.21, 0.67, and 89.7%, respectively. The general trends of greater deposition in the lungs than in the URT were similar in both studies, although much greater clearance to the esophagus and stomach was observed in this study. In comparison, for 9.65-µm particles, the fractions deposited in the lungs, trachea, URT, and GI tract were 0.44, 0.26, 74.8, and 24.5%, respectively (41). In this study, for the FFAG (10- to 15-µm particles), the fractions deposited in the lungs, trachea, URT, and GI tract were 1.4, 0.64, 36.5, and 61.5%, respectively. To our knowledge, this study represents the first description of the deposition of an aerosol comprising predominantly 12-µm particles within the murine respiratory tract. Differences between the two studies may in part be due to intraspecies anatomical and physiological differences; for example, BALB/c and B6C3F1 mice possess significant differences in airway anatomy, correlating with differences in body length and chest circumference (38, 39). A significant portion of the deposited material was cleared to the GI tract, possibly via epithelial mucociliary clearance mechanisms in some areas of the nasal passages and the tracheobronchial system. In the murine model (strain C57BL/6), clearance rates from the mucociliary escalators are rapid, at 2.2 ± 0.45 mm min–1 and 1.3 ± 0.26 mm min–1 for the trachea and nasal passages, respectively (18). The quantity of deposited bacteria cleared to the GI tract raises the important question of whether an inhalational challenge could result in "dual" infection of both the respiratory region and the GI tract. This would depend on microbial survival of the physical defense mechanisms in the stomach and intestine, including pH extremes and enzymatic degradation.
In conclusion, the FFAG is capable of delivering large-particle aerosols in the region of 12 µm to the nasal passages of the murine model within BSL-3 containment. The aerosol generated differs significantly from that produced by the Collison nebulizer with respect to mass distribution, distribution of particulates within the particles, and deposition within the murine respiratory tract. Flow-focusing technology has the potential to be utilized for the investigation of aerobiological research topics within diverse areas of applied microbiology, such as agriculture, food processing, water, biodefense, and health care industries. The ability to preferentially deposit pathogens within different regions of the respiratory tract will permit analysis of disease kinetics with respect to particle size. Future work will utilize the two aerosol generators to characterize the infections generated by differential deposition of aerosolized bacteria to the lungs or URT.
Special thanks are reserved for the personnel involved in the animal husbandry related to completion of the experiments.
Published ahead of print on 22 August 2008. ![]()
|
|
|---|
an-Calvo, A. M. 1998. Generation of steady liquid microthreads and micron-sized monodisperse sprays in gas streams. Phys. Rev. Lett. 80:285-288.[CrossRef]
an-Calvo, A. M., and A. Barrero. 1999. A novel pneumatic technique to generate steady capillary microjets. J. Aerosol Sci. 30:117-125.[Medline]
an-Calvo. 2005. Flow focusing: a versatile technology to produce size-controlled and specific-morphology microparticles. Small 1:688-692.[CrossRef][Medline]This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||