The D value of commercial biological indicator spore
strips using Bacillus stearothermophilus ATCC 7953 was
increased by higher calcium concentrations in assay media. The calcium
concentration in assay media varied among the manufacturers. The
calcium concentration in assay media is an important factor to consider
to minimize the variation of D value.